Comparison of LC-MS3 and LC-MRM Methods for Quantifying Amantadine and Its Application in Therapeutic Amantadine Monitoring in Human Plasma.
Sun, Qiang; Cao, Haiwei; Liu, Yong; Li, Yanyan; Huang, Jing.
Afiliação
  • Sun Q; Department of Laboratory Medicine, The First Hospital of Jilin University, Changchun 130061, China.
  • Cao H; Department of Laboratory Medicine, The First Hospital of Jilin University, Changchun 130061, China.
  • Liu Y; Genetic Diagnosis Center, The First Hospital of Jilin University, Changchun 130061, China.
  • Li Y; Department of Laboratory Medicine, The First Hospital of Jilin University, Changchun 130061, China.
  • Huang J; Department of Laboratory Medicine, The First Hospital of Jilin University, Changchun 130061, China.
Molecules ; 27(21)2022 Nov 07.
Article em En | MEDLINE | ID: mdl-36364446
ABSTRACT
A simple sample preprocessing method was developed for the quantitative determination of amantadine (AMT) in human plasma by liquid chromatography-tandem mass spectrometry cubed (LC-MS3). The LC-MS3 system comprised a Shimadzu Exion LC-20AD HPLC pump coupled with a QTRAP 5500 mass spectrometer. First, the plasma samples were pretreated using acetonitrile as the extracting solution to precipitate protein. Next, amantadine and amantadine-d15 (AMT-d15) were separated on an Agilent Poroshell 120 SB-C18 column (4.6 mm × 50 mm, 2.7 µm) using isocratic elution with solvent A (70% 0.1% formic acid) and solvent B (30% acetonitrile) at a flow rate of 0.8 mL/min. The total run time for each sample was 3 min. The system used triple-stage fragmentation transitions at m/z 152.2→135.3→107.4 for AMT quantification in the positive ion mode and m/z 167.0→150.3→118.1 for AMT-d15 quantification. The LC-MS3 assay was linear (r > 0.995) with a concentration range of 50−1500 ng/mL. The lower limit of quantification (LLOQ) was 50 ng/mL, and the intra-day and inter-day accuracies and precisions were less than 8.0% at all concentrations. In addition, the recoveries and matrix effect for AMT in human plasma were within acceptable limits. In terms of stability, AMT had no significant degradation under all conditions. All the results met the requirements of the guidelines of the Food and Drug Administration (FDA) for biological method validation. The novelty of the MS3 assay was that it presented a methodology with higher selectivity and sensitivity. This method was successfully applied to 44 human plasma samples, and the obtained quantitative results were compared with another liquid chromatography-multiple reaction monitoring (LC-MRM) method. The Passing-Bablok regression coefficients and Bland-Altman plot revealed no difference between the LC-MS3 and LC-MRM methods, implying that the developed LC-MS3 method is a reliable and accurate assay for AMT determination in human plasma. These results are also a proof of concept for determining chemicals in biological samples by the LC-MS3 strategy.