Your browser doesn't support javascript.
loading
Gas chromatography coupled to time-of-flight mass spectrometry using parallel electron and chemical ionization with permeation tube facilitated reagent ion control for material emission analysis.
Bräkling, Steffen; Hinterleitner, Christina; Cappellin, Luca; Vetter, Marleen; Mayer, Ingo; Benter, Thorsten; Klee, Sonja; Kersten, Hendrik.
Afiliação
  • Bräkling S; TOFWERK, Thun, Switzerland.
  • Hinterleitner C; Department of Physical and Theoretical Chemistry, University of Wuppertal, Wuppertal, Germany.
  • Cappellin L; Architecture, Wood and Civil Engineering, Bern University of Applied Science, Biel, Switzerland.
  • Vetter M; TOFWERK, Thun, Switzerland.
  • Mayer I; Department of Chemical Science, University of Padua, Padua, Italy.
  • Benter T; TOFWERK, Thun, Switzerland.
  • Klee S; Architecture, Wood and Civil Engineering, Bern University of Applied Science, Biel, Switzerland.
  • Kersten H; Department of Physical and Theoretical Chemistry, University of Wuppertal, Wuppertal, Germany.
Rapid Commun Mass Spectrom ; 37(7): e9461, 2023 Apr 15.
Article em En | MEDLINE | ID: mdl-36565273
ABSTRACT
RATIONALE Volatile organic compounds (VOCs) emitted by an artificial leather part for car interiors are determined using GC-MS (gas chromatograph coupled to a mass spectrometer) using simultaneous electron and chemical ionization (EI&CI). A device for swift reagent ion switching in CI mode between consecutive runs is presented.

METHODS:

VOCs emitted from the investigated material were sampled onto Tenax® absorption tubes using micro emission chambers and subsequently injected into the GC through thermal desorption. The detector was a time-of-flight mass spectrometer (TOFMS) simultaneously operating in EI and CI modes during a single chromatographic run. A custom permeation tube setup allowed for swift selection between various reagent ions in CI mode, e.g., [N2 H]+ , [H3 O]+ , [(H2 O)2 H]+ , and [NH4 ]+ .

RESULTS:

Different reagent ions are swiftly selectable between single GC runs without hardware changes. Differences in precursor ion survival yields and the selectivity of the various reactants were carefully assessed. Several examples for the improved identification of unknown compounds with the available complementary and comprehensive EI&CI data set are demonstrated for a relevant material emission application.

CONCLUSION:

The presented technique provides additional value to the standard GC-EI/MS procedure commonly used for material emission characterization. It allows for a non-targeted analysis approach with moderate analysis time.