Your browser doesn't support javascript.
loading
Elimination of lipaemic interference by high-speed centrifugation.
Solé-Enrech, Gemma; Cano-Corres, Ruth; Aparicio-Calvente, Maria Isabel; Spataro, Nino.
Afiliação
  • Solé-Enrech G; Clinical laboratory, Biochemistry Department, Parc Taulí Research and Innovation Institute Foundation (I3PT), Sabadell, Spain.
  • Cano-Corres R; Clinical laboratory, Biochemistry Department, Parc Taulí Research and Innovation Institute Foundation (I3PT), Sabadell, Spain.
  • Aparicio-Calvente MI; Clinical laboratory, Biochemistry Department, Parc Taulí Research and Innovation Institute Foundation (I3PT), Sabadell, Spain.
  • Spataro N; Clinical laboratory, Biochemistry Department, Parc Taulí Research and Innovation Institute Foundation (I3PT), Sabadell, Spain.
Biochem Med (Zagreb) ; 33(1): 010703, 2023 Feb 15.
Article em En | MEDLINE | ID: mdl-36627977
ABSTRACT

Introduction:

In order to deliver high quality results, detection and elimination of possible analytical interferences, such as lipaemia, is crucial. The aim of this study is to evaluate the efficacy of high-speed centrifugation in eliminating lipaemic interference and to define own lipaemic index (LI) for the studied biochemical analytes. Materials and

methods:

Evaluated analytes were albumin, alkaline phosphatase, alanine-aminotransferase (ALT), aspartate-aminotransferase (AST), calcium, creatinine, gamma-glutamyltransferase (GGT), glucose, phosphates, total proteins, urea and total bilirubin. Those analytes and LIs have been analysed in duplicate in the Roche Diagnostics-c8000 analyser in samples centrifuged at 3000 rpm/10 minutes in the SL16 (Thermo Scientific, Waltham, USA) centrifuge and according to an own high-speed centrifugation protocol (12,900 rpm/15 minutes) in the MicroCL17R (Thermo Scientific, Waltham, USA) centrifuge. Lipaemia has been measured in each sample. The efficiency of high-speed centrifugation is verified by the Wilcoxon test (P < 0.05). In cases where significant differences are observed, our own LI is calculated. For ALT and AST, it is verified by McNemar test (P < 0.05). For creatinine, both Wilcoxon and McNemar test were applied.

Results:

There were statistically significant differences in analyte concentration before and after high-speed centrifugation for albumin, creatinine, GGT, glucose, phosphates, urea and total bilirrubin. Own LI is calculated. McNemar test shows statistically significant diferences in the proportion of delivered results before and after high-speed centrifugation in ALT, AST and creatinine.

Conclusions:

This study confirms the efficacy of high-speed centrifugation protocol for all the considered analytes, excepting calcium, alkaline phosphatase and total proteins.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cálcio / Hiperlipidemias Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: Biochem Med (Zagreb) Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Espanha

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cálcio / Hiperlipidemias Tipo de estudo: Guideline Limite: Humans Idioma: En Revista: Biochem Med (Zagreb) Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Espanha