Comparison of Phacoemulsification Grooving Efficiency in Longitudinal vs Transversal Handpieces.
Wilkinson, Samuel W; Ungricht, Emilie L; West, William B; Harris, Jacob T; Zaugg, Brian; Olson, Randall J; Pettey, Jeff H.
Afiliação
  • Wilkinson SW; Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, UT, USA.
  • Ungricht EL; Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, UT, USA.
  • West WB; University of Utah School of Medicine, Salt Lake City, UT, USA.
  • Harris JT; Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, UT, USA.
  • Zaugg B; University of Utah School of Medicine, Salt Lake City, UT, USA.
  • Olson RJ; Department of Ophthalmology and Visual Sciences, John A. Moran Eye Center, University of Utah, Salt Lake City, UT, USA.
  • Pettey JH; University of Utah School of Medicine, Salt Lake City, UT, USA.
Clin Ophthalmol ; 17: 191-195, 2023.
Article em En | MEDLINE | ID: mdl-36660311
Purpose: To determine the difference, if any, in grooving efficiency at various settings on the Whitestar Signature Pro phacoemulsification (phaco) system. Methods: Cataractous lenses were simulated by exposing porcine lenses to formalin for 2 hours. A total of 120 lenses were analyzed at various power settings on both longitudinal and transversal handpieces. Twenty trials each were performed with power set to 25%, 50%, and 75% on both handpieces. A Whitestar Phaco Handpiece System was used to provide longitudinal power, and a Whitestar Signature Ellipsis Handpiece provided transversal power. Lenses were placed within a plastic chamber and grooved by an investigator blinded to settings. A second investigator recorded times and adjusted settings. The Whitestar Signature Pro phaco system was used for grooving. Results: There was no significant difference in grooving times between the longitudinal and transverse handpieces at any power setting (P > 0.05). There was a significant decrease in grooving times when comparing the 25% power setting with the 75% power setting for the transversal handpiece (P=0.021). Conclusion: Both longitudinal and transversal handpieces on the Whitestar Signature Pro phaco system produce similar results to one another at each power setting. There is a general trend toward shorter grooving times, reflecting greater efficiency, at higher power settings. Grooving efficiency on the transversal handpiece may be more affected by changes in the power settings as compared with the longitudinal settings.