Unusual atrial pacing induced by type II far-field p-wave sensing in a patient with complete atrioventricular block.
Pacing Clin Electrophysiol
; 47(3): 433-436, 2024 03.
Article
em En
| MEDLINE
| ID: mdl-37221929
An 88-year-old Japanese woman underwent DDD pacemaker (MicroPort KORA 250 DR, V lead: VEGA R52) implantation for complete atrioventricular block. A 12-lead electrocardiogram for a routine examination showed atrial pacing within the intrinsic P wave, followed by inhibition of ventricular pacing. Pacemaker interrogation revealed no abnormalities in the basic parameters; however, ventricular pacing was inhibited by far-field sensing of intrinsic atrial waves before atrial events; type II far-field P-wave sensing. As a result, unusual atrial pacing occurred due to the pause suppression algorithm, which is the one of the functions that prevent atrial fibrillation development.
Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Marca-Passo Artificial
/
Bloqueio Atrioventricular
Limite:
Aged80
/
Female
/
Humans
Idioma:
En
Revista:
Pacing Clin Electrophysiol
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Japão