Your browser doesn't support javascript.
loading
Determining the resolution of a tracer for magnetic particle imaging by means of magnetic particle spectroscopy.
Remmo, Amani; Wiekhorst, Frank; Kosch, Olaf; Lyer, Stefan; Unterweger, Harald; Kratz, Harald; Löwa, Norbert.
Afiliação
  • Remmo A; Physikalisch-Technische Bundesanstalt Berlin, Metrology for Magnetic Nanoparticles Abbestr. 2-12 10587 Berlin Germany amani.remmo@ptb.de.
  • Wiekhorst F; Physikalisch-Technische Bundesanstalt Berlin, Metrology for Magnetic Nanoparticles Abbestr. 2-12 10587 Berlin Germany amani.remmo@ptb.de.
  • Kosch O; Physikalisch-Technische Bundesanstalt Berlin, Metrology for Magnetic Nanoparticles Abbestr. 2-12 10587 Berlin Germany amani.remmo@ptb.de.
  • Lyer S; Department of Otorhinolaryngology, Head and Neck Surgery, Section of Experimental Oncology and Nanomedicine (SEON), Professorship for AI-Controlled Nanomaterials, Universitätsklinikum Erlangen Erlangen Germany.
  • Unterweger H; Department of Otorhinolaryngology, Head and Neck Surgery, Section of Experimental Oncology and Nanomedicine (SEON), Professorship for AI-Controlled Nanomaterials, Universitätsklinikum Erlangen Erlangen Germany.
  • Kratz H; Charité-Universitätsmedizin Berlin, Clinic for Radiology Charitéplatz 1 10117 Berlin Germany.
  • Löwa N; Physikalisch-Technische Bundesanstalt Berlin, Metrology for Magnetic Nanoparticles Abbestr. 2-12 10587 Berlin Germany amani.remmo@ptb.de.
RSC Adv ; 13(23): 15730-15736, 2023 May 22.
Article em En | MEDLINE | ID: mdl-37235104
ABSTRACT
Magnetic particle imaging (MPI) is an imaging modality to quantitatively determine the three-dimensional distribution of magnetic nanoparticles (MNPs) administered as a tracer into a biological system. Magnetic particle spectroscopy (MPS) is the zero-dimensional MPI counterpart without spatial coding but with much higher sensitivity. Generally, MPS is employed to qualitatively evaluate the MPI capability of tracer systems from the measured specific harmonic spectra. Here, we investigated the correlation of three characteristic MPS parameters with the achievable MPI resolution from a recently introduced procedure based on a two-voxel-analysis of data taken from the system function acquisition that is mandatory in Lissajous scanning MPI. We evaluated nine different tracer systems and determined their MPI capability and resolution from MPS measurements and compared the results with MPI phantom measurements.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: RSC Adv Ano de publicação: 2023 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: RSC Adv Ano de publicação: 2023 Tipo de documento: Article