A technique to measure tidal volume during noninvasive respiratory support by continuous-flow helmet CPAP.
Coppadoro, Andrea; Bellani, Giacomo; Foti, Giuseppe.
Afiliação
  • Coppadoro A; Department of Anesthesia and Intensive care, San Gerardo Hospital, Fondazione IRCCS San Gerardo dei Tintori, Monza, Italy.
  • Bellani G; Centre for Medical Sciences - CISMed, University of Trento, Trento, Italy.
  • Foti G; Department of Anesthesia and Intensive Care, Santa Chiara Regional Hospital, APSS Trento, Trento, Italy.
J Clin Monit Comput ; 37(6): 1473-1479, 2023 12.
Article em En | MEDLINE | ID: mdl-37329389
ABSTRACT

PURPOSE:

The coronavirus disease 2019 (COVID-19) pandemic has promoted the use of helmet continuous positive airway pressure (CPAP) for noninvasive respiratory support in hypoxic respiratory failure patients, despite the lack of tidal volume monitoring. We evaluated a novel technique designed to measure tidal volume during noninvasive continuous-flow helmet CPAP.

METHODS:

A bench model of spontaneously breathing patients undergoing helmet CPAP therapy (three positive end-expiratory pressure [PEEP] levels) at different levels of respiratory distress was used to compare measured and reference tidal volumes. Tidal volume measurement by the novel technique was based on helmet outflow-trace analysis. Helmet inflow was increased from 60 to 75 and 90 L/min to match the patient's peak inspiratory flow; an additional subset of tests was conducted under the condition of purposely insufficient inflow (i.e., high respiratory distress and 60 L/min inflow).

RESULTS:

The tidal volumes examined herein ranged from 250 to 910 mL. The Bland‒Altman analysis showed a bias of -3.2 ± 29.3 mL for measured tidal volumes compared to the reference, corresponding to an average relative error of -1 ± 4.4%. Tidal volume underestimation correlated with respiratory rate (rho = .411, p = .004) but not with peak inspiratory flow, distress, or PEEP. When the helmet inflow was maintained purposely low, tidal volume underestimation occurred (bias - 93.3 ± 83.9 mL), corresponding to an error of -14.8 ± 6.3%.

CONCLUSION:

Tidal volume measurement is feasible and accurate during bench continuous-flow helmet CPAP therapy by the analysis of the outflow signal, provided that helmet inflow is adequate to match the patient's inspiratory efforts. Insufficient inflow resulted in tidal volume underestimation. In vivo data are needed to confirm these findings.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Síndrome do Desconforto Respiratório / Insuficiência Respiratória Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: J Clin Monit Comput Assunto da revista: INFORMATICA MEDICA / MEDICINA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Itália

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Síndrome do Desconforto Respiratório / Insuficiência Respiratória Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: J Clin Monit Comput Assunto da revista: INFORMATICA MEDICA / MEDICINA Ano de publicação: 2023 Tipo de documento: Article País de afiliação: Itália