Your browser doesn't support javascript.
loading
Mobile superconducting strip photon detection system with efficiency over 70% at a 1550 nm wavelength.
Opt Express ; 31(19): 30650-30657, 2023 Sep 11.
Article em En | MEDLINE | ID: mdl-37710604
We developed a mobile superconducting strip photon detector (SSPD) system operated in a liquid-helium Dewar. By adopting highly disordered NbTiN thin films, we successfully enhanced the detection performance of superconducting strips at higher operation temperatures and realized SSPDs with nearly saturated detection efficiency at 4.2 K. Then we customized a compact liquid-helium Dewar and a battery-based electronic module to minimize the SSPD system. A mobile SSPD system was integrated, which showed a system detection efficiency of 72% for a 1550 nm wavelength with a dark count rate of 200 cps and a timing jitter of 67.2 ps. The system has a weight of 40 kg and a power consumption of 500 mW, which can work continuously for 20 hours. The metrics can be further optimized in accordance with the various practical application platforms, such as aircraft, drones, etc.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Tipo de estudo: Diagnostic_studies Idioma: En Revista: Opt Express Assunto da revista: OFTALMOLOGIA Ano de publicação: 2023 Tipo de documento: Article