Biomechanical comparison of suture bridge rotator cuff repair with and without dermal allograft pledgets.
Tashjian, Robert Z; Chalmers, Peter N; Joyce, Christopher D; Asghar, Elise B; Henninger, Heath B.
Afiliação
  • Tashjian RZ; Department of Orthopaedic Surgery, University of Utah, Salt Lake City, UT, USA.
  • Chalmers PN; Department of Orthopaedic Surgery, University of Utah, Salt Lake City, UT, USA. Electronic address: p.n.chalmers@gmail.com.
  • Joyce CD; Department of Orthopaedic Surgery, University of Utah, Salt Lake City, UT, USA.
  • Asghar EB; Redwood Orthopaedic Surgery Associates, Santa Rosa, CA, USA.
  • Henninger HB; Department of Orthopaedic Surgery, University of Utah, Salt Lake City, UT, USA.
J Shoulder Elbow Surg ; 33(6): 1360-1365, 2024 Jun.
Article em En | MEDLINE | ID: mdl-38122892
ABSTRACT

BACKGROUND:

One method to augment rotator cuff repair is to pass dermal allograft pledgets along the sutures that bridge from the medial to the lateral row. It remains unclear whether this augmentation method alters repair biomechanics.

METHODS:

This was a controlled laboratory study. After an a priori power analysis, 9 pairs of rotator cuffs underwent double-row suture bridge rotator cuff repair, half randomized to augmentation with dermal allograft pledgets passed along the suture bridge sutures. Repairs were then mounted on a material testing system and loaded cyclically 500 cycles to measure applied force and displacement. Repairs then underwent ultimate failure testing, and stiffness, ultimate failure force, and ultimate failure displacement were measured. Paired t tests were performed to compare between groups.

RESULTS:

There were no differences between groups in construct gapping with cyclic loading after 500 cycles (P = .885). There were no differences between the augmented and control groups in yield force (103.5 ± 5.0 vs. 101.4 ± 5.9 N, respectively, P = .183), stiffness (94.2 ± 13.9 vs. 90.9 ± 13.8, P = .585), or ultimate failure force (255.3 ± 65.8 vs. 285.3 ± 83.2, P = .315). There were no differences between groups in failure modes, with most specimens failing by cuff tissue tearing within or medial to the construct.

CONCLUSION:

The addition of dermal allograft pledgets does not positively or negatively influence the time-zero biomechanical characteristics of double-row suture bridge rotator cuff repair.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas de Sutura / Lesões do Manguito Rotador Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Shoulder Elbow Surg Assunto da revista: ORTOPEDIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Técnicas de Sutura / Lesões do Manguito Rotador Limite: Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Shoulder Elbow Surg Assunto da revista: ORTOPEDIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos