Your browser doesn't support javascript.
loading
Multibody dynamic modeling of the behavior of flexible instruments used in cervical cancer brachytherapy.
Straathof, Robin; Meijaard, Jaap P; van Vliet-Pérez, Sharline M; Kolkman-Deurloo, Inger-Karine K; Nout, Remi A; Heijmen, Ben J M; Wauben, Linda S G L; Dankelman, Jenny; van de Berg, Nick J.
Afiliação
  • Straathof R; Department of BioMechanical Engineering, Delft University of Technology, Delft, the Netherlands.
  • Meijaard JP; Department of Radiotherapy, Erasmus MC Cancer Institute, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • van Vliet-Pérez SM; Department of Precision and Microsystems Engineering, Delft University of Technology, Delft, the Netherlands.
  • Kolkman-Deurloo IK; Department of BioMechanical Engineering, Delft University of Technology, Delft, the Netherlands.
  • Nout RA; Department of Radiotherapy, Erasmus MC Cancer Institute, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • Heijmen BJM; Department of Radiotherapy, Erasmus MC Cancer Institute, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • Wauben LSGL; Department of Radiotherapy, Erasmus MC Cancer Institute, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • Dankelman J; Department of Radiotherapy, Erasmus MC Cancer Institute, University Medical Center Rotterdam, Rotterdam, the Netherlands.
  • van de Berg NJ; Department of BioMechanical Engineering, Delft University of Technology, Delft, the Netherlands.
Med Phys ; 51(5): 3698-3710, 2024 May.
Article em En | MEDLINE | ID: mdl-38226798
ABSTRACT

BACKGROUND:

The steep radiation dose gradients in cervical cancer brachytherapy (BT) necessitate a thorough understanding of the behavior of afterloader source cables or needles in the curved channels of (patient-tailored) applicators.

PURPOSE:

The purpose of this study is to develop and validate computer models to simulate (1) BT source positions, and (2) insertion forces of needles in curved applicator channels. The methodology presented can be used to improve the knowledge of instrument behavior in current applicators and aid the development of novel (3D-printed) BT applicators.

METHODS:

For the computer models, BT instruments were discretized in finite elements. Simulations were performed in SPACAR by formulating nodal contact force and motion input models and specifying the instruments' kinematic and dynamic properties. To evaluate the source cable model, simulated source paths in ring applicators were compared with manufacturer-measured source paths. The impact of discrepancies on the dosimetry was estimated for standard plans. To validate needle models, simulated needle insertion forces in curved channels with varying curvature, torsion, and clearance, were compared with force measurements in dedicated 3D-printed templates.

RESULTS:

Comparison of simulated with manufacturer-measured source positions showed 0.5-1.2 mm median and <2.0 mm maximum differences, in all but one applicator geometry. The resulting maximum relative dose differences at the lateral surface and at 5 mm depth were 5.5% and 4.7%, respectively. Simulated insertion forces for BT needles in curved channels accurately resembled the forces experimentally obtained by including experimental uncertainties in the simulation.

CONCLUSION:

The models developed can accurately predict source positions and insertion forces in BT applicators. Insights from these models can aid novel applicator design with improved motion and force transmission of BT instruments, and contribute to the estimation of overall treatment precision. The methodology presented can be extended to study other applicator geometries, flexible instruments, and afterloading systems.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Braquiterapia / Neoplasias do Colo do Útero Tipo de estudo: Prognostic_studies Limite: Female / Humans Idioma: En Revista: Med Phys Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Holanda

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Braquiterapia / Neoplasias do Colo do Útero Tipo de estudo: Prognostic_studies Limite: Female / Humans Idioma: En Revista: Med Phys Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Holanda