Your browser doesn't support javascript.
loading
Hyperthermic pressurized intraperitoneal aerosol drug delivery system in a large animal model: a feasibility and safety study.
Min, Sa-Hong; Yoo, Mira; Hwang, Duyeong; Lee, Eunju; Kang, So Hyun; Lee, Sangjun; Won, Yongjoon; Park, Young Suk; Ahn, Sang-Hoon; Kim, Hyung-Ho.
Afiliação
  • Min SH; Department of Gastrointestinal Surgery, Asan Medical Center, Seoul, Korea.
  • Yoo M; Department of Surgery, Seoul National University Bundang Hospital, Seongnam, Korea.
  • Hwang D; Department of Surgery, Seoul National University Bundang Hospital, Seongnam, Korea.
  • Lee E; Department of Surgery, Chung-Ang University Gwangmyeong Hospital, Gwangmyeong, Korea.
  • Kang SH; Department of Surgery, Seoul National University Bundang Hospital, Seongnam, Korea.
  • Lee S; Department of Surgery, Kyung Hee University Hospital At Gangdong, Seoul, Korea.
  • Won Y; Department of Surgery, Seongnam Citizens Medical Center, Seongnam, Korea.
  • Park YS; Department of Surgery, Seoul National University Bundang Hospital, Seongnam, Korea.
  • Ahn SH; Department of Surgery, Seoul National University Bundang Hospital, Seongnam, Korea.
  • Kim HH; Department of Surgery, Seoul National University Bundang Hospital, Seongnam, Korea. lapakh2@cauhs.or.kr.
Surg Endosc ; 38(4): 2062-2069, 2024 Apr.
Article em En | MEDLINE | ID: mdl-38429574
ABSTRACT

BACKGROUND:

We developed a novel drug delivery system called hyperthermic pressurized intraperitoneal aerosol chemotherapy (HPIPAC) that hybridized Hyperthermic intraperitoneal chemotherapy (HIPEC) and pressurized intraperitoneal aerosol chemotherapy (PIPAC). The present study aims to assess the feasibility and safety of HPIPAC system in a large animal survival model.

METHODS:

Eleven pigs (eight non-survival models and three survival models) were used in the experiment. The heat module in the HPIPAC controller circulates hyperthermic CO2 in a closed-loop circuit and creates gas-based dry intraperitoneal hyperthermia. Three 12 mm trocars were placed on the abdomen. The afferent CO2 tube wound with heat generating coil was inserted into a trocar, and the efferent tube was inserted into another trocar. Heated CO2 was insufflated and circulated in a closed circuit until the intra-abdominal and peritoneal surface temperature reached 42 °C. 100 ml of 5% dextrose in water was nebulized for 5 min and the closed-loop circulation was resumed for 60 min at 42 °C. Tissue biopsies were taken from several sites from the pigs in the survival model.

RESULTS:

The average change in core temperature of the pigs was 2.5 ± 0.08 °C. All three pigs displayed no signs of distress, and their vital signs remained stable, with no changes in their diet. In autopsy, inflammatory and fibrotic responses at the biopsy sites were observed without serious pathologic findings.

CONCLUSIONS:

We successfully proved the feasibility and safety of our novel HPIPAC system in an in-vivo swine survival model.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias Peritoneais Limite: Animals Idioma: En Revista: Surg Endosc Assunto da revista: DIAGNOSTICO POR IMAGEM / GASTROENTEROLOGIA Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Neoplasias Peritoneais Limite: Animals Idioma: En Revista: Surg Endosc Assunto da revista: DIAGNOSTICO POR IMAGEM / GASTROENTEROLOGIA Ano de publicação: 2024 Tipo de documento: Article