MR-linac MLC positioning QA by digitally stitching dual double-exposed films.
J Appl Clin Med Phys
; 25(7): e14325, 2024 Jul.
Article
em En
| MEDLINE
| ID: mdl-38467039
ABSTRACT
PURPOSE:
The picket fence (PF) test is highly recommended for multi-leaf collimator (MLC) quality assurance. However, since the electronic portal imaging device (EPID) on the Elekta Unity only covers a small area, it is not feasible to perform the PF test for the entire MLC. Here, we propose a technique for the PF test by stitching two double-exposed films.METHODS:
Two EBT3 films were used to encompass the entire MLC, with each one covering one half of the area. Two fields were employed to apply double exposure a PF pattern consisting of 11 2 mm wide pickets and a 2.84 cm x 22 cm open field. The edges of the open field defined by the diaphragms were used to correct film rotation as well as align them horizontally. The PF pattern was also measured with the EPID where the pickets were used to align the films vertically. Individual leaf positions were detected on the merged film for quantitative analysis. Various MLC positioning errors were introduced to evaluate the technique's sensitivity.RESULTS:
The merged films covered 72 leaf pairs properly (four leaf pairs on both sides were outside the treatment couch). With the EPID, the leaf positioning accuracy was -0.02 ± 0.07 mm (maximum 0.29 mm) and the picket width variation was 0.00 ± 0.03 mm (maximum 0.11 mm); with the films, the position accuracy and width variation were -0.03 ± 0.13 mm (maximum 0.80 mm) and 0.00 ± 0.13 mm (maximum 0.74 mm), respectively. The EPID was able to detect errors of 0.5 mm or above with submillimeter accuracy; the films were only able to detect errors > 1.0 mm.CONCLUSION:
We developed a quantitative technique for the PF test on the Elekta Unity. The merged films covered nearly the entire MLC leaf banks. The technique exhibited clinically acceptable accuracy and sensitivity to MLC positioning errors.Palavras-chave
Texto completo:
1
Coleções:
01-internacional
Base de dados:
MEDLINE
Assunto principal:
Aceleradores de Partículas
/
Garantia da Qualidade dos Cuidados de Saúde
/
Dosagem Radioterapêutica
/
Planejamento da Radioterapia Assistida por Computador
/
Radioterapia de Intensidade Modulada
Limite:
Humans
Idioma:
En
Revista:
J Appl Clin Med Phys
Assunto da revista:
BIOFISICA
Ano de publicação:
2024
Tipo de documento:
Article
País de afiliação:
Estados Unidos