Your browser doesn't support javascript.
loading
Biomechanical behavior of implant retained prostheses in the posterior maxilla using different materials: a finite element study.
Aboelfadl, Ahmad; Keilig, Ludger; Ebeid, Kamal; Ahmed, Mohamed Abdel Moniem; Nouh, Ingy; Refaie, Ashraf; Bourauel, Christoph.
Afiliação
  • Aboelfadl A; Department of Fixed Prosthodontics, Faculty of Dentistry, Ain Shams University, Cairo, Egypt. a.aboelfadl@gmail.com.
  • Keilig L; Oral Technology, Dental School, University Hospital Bonn, Bonn, Germany. a.aboelfadl@gmail.com.
  • Ebeid K; Oral Technology, Dental School, University Hospital Bonn, Bonn, Germany.
  • Ahmed MAM; Department of Dental Prosthetics, Propaedeutics and Material Science, Dental School, University Hospital Bonn, Bonn, Germany.
  • Nouh I; Department of Fixed Prosthodontics, Faculty of Dentistry, Ain Shams University, Cairo, Egypt.
  • Refaie A; Department of Fixed Prosthodontics, Faculty of Dentistry, Suez University, Suez, Egypt.
  • Bourauel C; Department of Fixed Prosthodontics, Faculty of Dentistry, Ain Shams University, Cairo, Egypt.
BMC Oral Health ; 24(1): 455, 2024 Apr 15.
Article em En | MEDLINE | ID: mdl-38622680
ABSTRACT

BACKGROUND:

The aim of this study is to evaluate the biomechanical behavior of the mesial and distal off-axial extensions of implant-retained prostheses in the posterior maxilla with different prosthetic materials using finite element analysis (FEA).

METHODS:

Three dimensional (3D) finite element models with three implant configurations and prosthetic designs (fixed-fixed, mesial cantilever, and distal cantilever) were designed and modelled depending upon cone beam computed tomography (CBCT) images of an intact maxilla of an anonymous patient. Implant prostheses with two materials; Monolithic zirconia (Zr) and polyetherketoneketone (PEKK) were also modeled .The 3D modeling software Mimics Innovation Suite (Mimics 14.0 / 3-matic 7.01; Materialise, Leuven, Belgium) was used. All the models were imported into the FE package Marc/Mentat (ver. 2015; MSC Software, Los Angeles, Calif). Then, individual models were subjected to separate axial loads of 300 N. Von mises stress values were computed for the prostheses, implants, and bone under axial loading.

RESULTS:

The highest von Mises stresses in implant (111.6 MPa) and bone (100.0 MPa) were recorded in distal cantilever model with PEKK material, while the lowest values in implant (48.9 MPa) and bone (19.6 MPa) were displayed in fixed fixed model with zirconia material. The distal cantilever model with zirconia material yielded the most elevated levels of von Mises stresses within the prosthesis (105 MPa), while the least stresses in prosthesis (35.4 MPa) were recorded in fixed fixed models with PEKK material.

CONCLUSIONS:

In the light of this study, the combination of fixed fixed implant prosthesis without cantilever using a rigid zirconia material exhibits better biomechanical behavior and stress distribution around bone and implants. As a prosthetic material, low elastic modulus PEKK transmitted more stress to implants and surrounding bone especially with distal cantilever.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Zircônio / Implantes Dentários Limite: Humans Idioma: En Revista: BMC Oral Health Assunto da revista: ODONTOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Egito

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Zircônio / Implantes Dentários Limite: Humans Idioma: En Revista: BMC Oral Health Assunto da revista: ODONTOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Egito