Your browser doesn't support javascript.
loading
Accuracy of the novel digital non-cross-arch surgical guides with integration of tooth undercut retention and screw-bone support for implant placement in mandibular free-end.
Wu, Qin; Lou, Yuxin; Sun, Jikui; Xie, Chenyang; Wu, Jiacheng; Yu, Haiyang.
Afiliação
  • Wu Q; State Key Laboratory of Oral Diseases & National Center for Stomatology & National Clinical Research Center for Oral Diseases, Department of Prosthodontics II, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd Section, Chengdu, Sichuan, 610041, China.
  • Lou Y; State Key Laboratory of Oral Diseases & National Center for Stomatology & National Clinical Research Center for Oral Diseases, Department of Dental Technology, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd Section, Chengdu, Sichuan, 610041, China.
  • Sun J; State Key Laboratory of Oral Diseases & National Center for Stomatology & National Clinical Research Center for Oral Diseases, Department of Prosthodontics II, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd Section, Chengdu, Sichuan, 610041, China.
  • Xie C; State Key Laboratory of Oral Diseases & National Center for Stomatology & National Clinical Research Center for Oral Diseases, Department of Dental Technology, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd Section, Chengdu, Sichuan, 610041, China.
  • Wu J; State Key Laboratory of Oral Diseases & National Center for Stomatology & National Clinical Research Center for Oral Diseases, Department of Dental Technology, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd Section, Chengdu, Sichuan, 610041, China.
  • Yu H; State Key Laboratory of Oral Diseases & National Center for Stomatology & National Clinical Research Center for Oral Diseases, Department of Prosthodontics II, West China Hospital of Stomatology, Sichuan University, 14 Renmin South Road, 3rd Section, Chengdu, Sichuan, 610041, China. YHYmecha
BMC Oral Health ; 24(1): 550, 2024 May 11.
Article em En | MEDLINE | ID: mdl-38734597
ABSTRACT

BACKGROUND:

Large cross-arch free-end surgical guides can obscure the visual field, compromising surgical accuracy due to insufficient stability at the free-end. This in vitro study aims to evaluate the accuracy of novel digital non-cross-arch surgical guides designed for implant placement at the mandibular free-end, incorporating tooth undercut retention and screw-bone support. MATERIALS AND

METHODS:

A mandibular dental model lacking left molars was utilized to fabricate unilateral (cross-arch) tooth-supported surgical guides (GT I, n = 20). Subsequently, two additional types of surgical guides were fabricated GT II (covering two teeth, n = 20) and GT III (covering three teeth, n = 20). These novel surgical guides were designed to utilize the undercut of the supporting teeth for retention and enhance stability with screw-bone support at the guide's free-end. Furthermore, 60 identical guiding blocks were assembled on the three types of surgical guides to facilitate the implants' insertion. On a phantom head, 120 implant replicas were placed at the Federal Dentaire Internationale (FDI) teeth positions #36 and #37 on the dental model, employing a combination of surgical guides and guiding blocks. To assess accuracy, planned and placed implant positions were compared using intraoral optical scanning. Discrepancies in angulation and linear deviations, including the coronal/apical 3D deviations, lateral deviation as well as depth deviation, were measured. Statistical analysis was performed using two-way ANOVA and Bonferroni test (α = 0.05).

RESULTS:

GT I exhibited significantly largest discrepancies, including angular and linear deviations at the crest and apex at every implant site. Especially in depth, at implant site #36, the mean deviation value of GT I (0.27 ± 0.13 mm) was twice as large as GT III (0.13 ± 0.07 mm), and almost twice as large as GT II (0.14 ± 0.08 mm). However, at implant site #37, this deviation increased to almost a five-fold relationship between GT I (0.63 ± 0.12 mm) and II (0.14 ± 0.09 mm), as well as between GT I and III (0.13 ± 0.09 mm). No significant discrepancies existed between the novel surgical guides at either implant site #36 or #37.

CONCLUSION:

This study provides a practical protocol for enhancing accuracy of implant placement and reducing the size of free-end surgical guides used at mandibular molar sites.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Parafusos Ósseos / Cirurgia Assistida por Computador / Modelos Dentários / Mandíbula Limite: Humans Idioma: En Revista: BMC Oral Health Assunto da revista: ODONTOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Parafusos Ósseos / Cirurgia Assistida por Computador / Modelos Dentários / Mandíbula Limite: Humans Idioma: En Revista: BMC Oral Health Assunto da revista: ODONTOLOGIA Ano de publicação: 2024 Tipo de documento: Article País de afiliação: China