Image quality of opportunistic breast examinations in photon-counting computed tomography: A phantom study.
Sawall, S; Baader, E; Wolf, J; Maier, J; Schlemmer, H-P; Schönberg, S O; Sechopoulos, I; Kachelrieß, M.
Afiliação
  • Sawall S; Division of X-Ray Imaging and CT, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg, Germany; Medical Faculty, Heidelberg University, Im Neuenheimer Feld 672, 69120 Heidelberg, Germany. Electronic address: stefan.sawall@dkfz.de.
  • Baader E; Division of X-Ray Imaging and CT, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg, Germany; Department of Physics and Astronomy, Ruprecht-Karls-University Heidelberg, Heidelberg, Germany.
  • Wolf J; Division of X-Ray Imaging and CT, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg, Germany.
  • Maier J; Division of X-Ray Imaging and CT, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg, Germany.
  • Schlemmer HP; Medical Faculty, Heidelberg University, Im Neuenheimer Feld 672, 69120 Heidelberg, Germany; Division of Radiology, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg, Germany.
  • Schönberg SO; Department of Radiology and Nuclear Medicine, University Medical Center Mannheim, Theodor-Kutzer-Ufer 1-3, 68167 Mannheim, Germany.
  • Sechopoulos I; Department of Medical Imaging, Radboud University Medical Center, Geert Grooteplein Zuid 10, 6525 GA Nijmegen, The Netherlands.
  • Kachelrieß M; Division of X-Ray Imaging and CT, German Cancer Research Center (DKFZ), Im Neuenheimer Feld 280, 69120 Heidelberg, Germany; Medical Faculty, Heidelberg University, Im Neuenheimer Feld 672, 69120 Heidelberg, Germany.
Phys Med ; 122: 103378, 2024 Jun.
Article em En | MEDLINE | ID: mdl-38797026
ABSTRACT

PURPOSE:

To compare the breast imaging performance of a clinical whole-body photon-counting CT (PCCT) to that of a dedicated breast CT (BCT) to determine the image quality of opportunistic breast examinations in clinical PCCT. MATERIALS AND

METHODS:

To quantify image quality for breast cancer applications, acquisitions of a breast phantom including representations of calcifications, fibers, and masses were performed using a clinical PCCT and a dedicated BCT. When imaging with the PCCT, the phantom was also combined with a thorax phantom to simulate realistic patient positioning, while only the breast phantom was imaged in the BCT. Images in BCT were acquired at 7.0 mGy (CTDI16cm) and using 2.6 mGy-25.0 mGy in the PCCT. Spatial resolution between the BCT and PCCT images was matched and data were reconstructed using the default methods of each system. The dose-normalized contrast-to-noise ratio (CNRD) of masses and the structural visibility of fibers and calcifications were evaluated as figures of merit for all reconstructions.

RESULTS:

CNRD between masses and background was 0.56 mGy-½, on average with BCT and varied between 0.39 mGy-½ to 1.46 mGy-½ with PCCT over all dose levels, phantom configurations, and reconstruction algorithms. Calcifications down to a size of 0.29 mm and fibers down to a size of 0.23 mm could be reliably identified in the images of both systems.

CONCLUSIONS:

Clinical PCCT provides an image quality superior to that obtained with BCT in terms of CNRD and allows for the identification of calcifications and fibers at comparable dose levels.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Fótons / Imagens de Fantasmas Limite: Female / Humans Idioma: En Revista: Phys Med / Phys. medica (Testo stamp.) / Physica medica Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Tomografia Computadorizada por Raios X / Fótons / Imagens de Fantasmas Limite: Female / Humans Idioma: En Revista: Phys Med / Phys. medica (Testo stamp.) / Physica medica Assunto da revista: BIOFISICA / BIOLOGIA / MEDICINA Ano de publicação: 2024 Tipo de documento: Article