Your browser doesn't support javascript.
loading
Wolfram-silicone implants as effective radiation shielding for ocular brachytherapy: dosimetric features and in vivo animal study on biocompatibility.
Sokolenko, Ekaterina A; Flühs, Dirk; Lalos, Fotis; Meyer, Peter; Fiorentzis, Miltiadis; Lindziute, Migle; Gemmecke, Justine; Berchner-Pfannschmidt, Utta; Hendgen-Cotta, Ulrike; Bechrakis, Nikolaos E; Tsimpaki, Theodora; Dubicanac, Marko; Wißmann, Andreas; Hilken, Gero.
Afiliação
  • Sokolenko EA; Institute of Ophthalmology, University Eye Hospital, Hannover Medical School, Hannover, Germany. Electronic address: sokolenko.ekaterina@mh-hannover.de.
  • Flühs D; Department of Radiotherapy, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.
  • Lalos F; MVZ Private Clinic of Ophthalmology, Altenessen GmbH, Essen, Germany.
  • Meyer P; Ophthalmology Department, University Hospital Basel, Basel, Switzerland.
  • Fiorentzis M; Department of Ophthalmology, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.
  • Lindziute M; Institute of Ophthalmology, University Eye Hospital, Hannover Medical School, Hannover, Germany.
  • Gemmecke J; Department of Physics, Technical University Dortmund, Dortmund, Germany.
  • Berchner-Pfannschmidt U; Department of Ophthalmology, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.
  • Hendgen-Cotta U; West German Heart and Vascular Center, Clinic for Cardiology and Angiology, University Hospital Essen, Essen, Germany.
  • Bechrakis NE; Department of Ophthalmology, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.
  • Tsimpaki T; Ophthalmology Department, University Hospital Basel, Basel, Switzerland.
  • Dubicanac M; Central Animal Laboratory, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.
  • Wißmann A; Central Animal Laboratory, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.
  • Hilken G; Central Animal Laboratory, University Hospital Essen, University of Duisburg-Essen, Essen, Germany.
Can J Ophthalmol ; 2024 May 28.
Article em En | MEDLINE | ID: mdl-38815956
ABSTRACT

OBJECTIVE:

To evaluate wolfram as a photon and beta absorber in the management of uveal melanoma with radiotherapy, examining its potential ocular adverse effects and physiologic tolerance using an in vivo rabbit ocular model.

METHODS:

A method of manufacturing implants from mixtures of wolfram and silicone was developed. Their shielding effect on the radiation of sources used in ocular brachytherapy was investigated by dosimetric measurement in an eye phantom as well as numerical simulations. Different wolfram implantation techniques, such as extraocular fixation of a wolfram-silicone implant (n = 1), vitrectomy with silicone oil and intravitreal injection of a wolfram-silicone oil suspension (n = 2), and concurrent attachment of a wolfram implant onto the sclera (n = 2), were tested to investigate the long-term effects of wolfram. A vitrectomy with silicone oil without wolfram implantation was carried out in 2 rabbits (n = 2), constituting the control group. The eyes were enucleated after 3 months for histologic analysis.

RESULTS:

Wolfram-silicone mixtures have been dosimetrically proven to be very effective radiation absorbers for use in ocular brachytherapy. Severe complications, such as endophthalmitis, secondary glaucoma, cornea decompensation, and vessel occlusion, were not documented in the tested rabbit eyes after the application of wolfram. Histologic examination of the bulbi after enucleation showed epiretinal gliosis without further pathologic findings in all eyes after vitrectomy.

CONCLUSIONS:

The results of this study show that wolfram and wolfram-silicone implants constitute a promising candidate as potential radiation shielding substrates.

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Can J Ophthalmol Ano de publicação: 2024 Tipo de documento: Article

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Idioma: En Revista: Can J Ophthalmol Ano de publicação: 2024 Tipo de documento: Article