Impact of reconstruction parameters on the accuracy of myocardial extracellular volume quantification on a first-generation, photon-counting detector CT.
Gnasso, Chiara; Pinos, Daniel; Schoepf, U Joseph; Vecsey-Nagy, Milan; Aquino, Gilberto J; Fink, Nicola; Zsarnoczay, Emese; Holtackers, Robert J; Stock, Jonathan; Suranyi, Pal; Varga-Szemes, Akos; Emrich, Tilman.
Afiliação
  • Gnasso C; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Dr, Charleston, SC, 29425, USA.
  • Pinos D; Experimental Imaging Center, IRCCS San Raffaele Scientific Institute, Via Olgettina 60, Milan, Italy.
  • Schoepf UJ; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Dr, Charleston, SC, 29425, USA.
  • Vecsey-Nagy M; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Dr, Charleston, SC, 29425, USA.
  • Aquino GJ; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Dr, Charleston, SC, 29425, USA.
  • Fink N; Heart and Vascular Center, Semmelweis University, Varosmajor Utca 68, Budapest, 1122, Hungary.
  • Zsarnoczay E; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Dr, Charleston, SC, 29425, USA.
  • Holtackers RJ; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Dr, Charleston, SC, 29425, USA.
  • Stock J; Department of Radiology, University Hospital, LMU Munich, Marchioninistr. 15, Munich, 81377, Germany.
  • Suranyi P; Division of Cardiovascular Imaging, Department of Radiology and Radiological Science, Medical University of South Carolina, 25 Courtenay Dr, Charleston, SC, 29425, USA.
  • Varga-Szemes A; MTA-SE Cardiovascular Imaging Research Group, Department of Radiology, Medical Imaging Centre, Semmelweis University, Ülloi Út 78, Budapest, 1082, Hungary.
  • Emrich T; Department of Radiology and Nuclear Medicine, Maastricht University Medical Centre, Maastricht, 6229 HX, The Netherlands.
Eur Radiol Exp ; 8(1): 70, 2024 Jun 19.
Article em En | MEDLINE | ID: mdl-38890175
ABSTRACT

BACKGROUND:

The potential role of cardiac computed tomography (CT) has increasingly been demonstrated for the assessment of diffuse myocardial fibrosis through the quantification of extracellular volume (ECV). Photon-counting detector (PCD)-CT technology may deliver more accurate ECV quantification compared to energy-integrating detector CT. We evaluated the impact of reconstruction settings on the accuracy of ECV quantification using PCD-CT, with magnetic resonance imaging (MRI)-based ECV as reference.

METHODS:

In this post hoc analysis, 27 patients (aged 53.1 ± 17.2 years (mean ± standard deviation); 14 women) underwent same-day cardiac PCD-CT and MRI. Late iodine CT scans were reconstructed with different quantum iterative reconstruction levels (QIR 1-4), slice thicknesses (0.4-8 mm), and virtual monoenergetic imaging levels (VMI, 40-90 keV); ECV was quantified for each reconstruction setting. Repeated measures ANOVA and t-test for pairwise comparisons, Bland-Altman plots, and Lin's concordance correlation coefficient (CCC) were used.

RESULTS:

ECV values did not differ significantly among QIR levels (p = 1.000). A significant difference was observed throughout different slice thicknesses, with 0.4 mm yielding the highest agreement with MRI-based ECV (CCC = 0.944); 45-keV VMI reconstructions showed the lowest mean bias (0.6, 95% confidence interval 0.1-1.4) compared to MRI. Using the most optimal reconstruction settings (QIR4. slice thickness 0.4 mm, VMI 45 keV), a 63% reduction in mean bias and a 6% increase in concordance with MRI-based ECV were achieved compared to standard settings (QIR3, slice thickness 1.5 mm; VMI 65 keV).

CONCLUSIONS:

The selection of appropriate reconstruction parameters improved the agreement between PCD-CT and MRI-based ECV. RELEVANCE STATEMENT Tailoring PCD-CT reconstruction parameters optimizes ECV quantification compared to MRI, potentially improving its clinical utility. KEY POINTS • CT is increasingly promising for myocardial tissue characterization, assessing focal and diffuse fibrosis via late iodine enhancement and ECV quantification, respectively. • PCD-CT offers superior performance over conventional CT, potentially improving ECV quantification and its agreement with MRI-based ECV. • Tailoring PCD-CT reconstruction parameters optimizes ECV quantification compared to MRI, potentially improving its clinical utility.
Assuntos
Palavras-chave

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Tomografia Computadorizada por Raios X / Miocárdio Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Eur Radiol Exp / European radiol. exp / European radiology experimental Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Imageamento por Ressonância Magnética / Tomografia Computadorizada por Raios X / Miocárdio Limite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: Eur Radiol Exp / European radiol. exp / European radiology experimental Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Estados Unidos