Your browser doesn't support javascript.
loading
An FDA-Validated, Self-Cleaning Liquid Chromatography-Mass Spectrometry System for Determining Small-Molecule Drugs and Metabolites in Organoid/Organ-on-Chip Medium.
Kogler, Stian; Pedersen, Gustav Mathingsdal; Martínez-Ramírez, Felipe; Aizenshtadt, Aleksandra; Busek, Mathias; Krauss, Stefan J K; Wilson, Steven Ray; Røberg-Larsen, Hanne.
Afiliação
  • Kogler S; Hybrid Technology Hub-Centre of Excellence, Institute of Basic Medical Sciences, Faculty of Medicine, University of Oslo, Oslo 0372, Norway.
  • Pedersen GM; Section for Chemical Life Sciences, Department of Chemistry, University of Oslo, Oslo NO-0315, Norway.
  • Martínez-Ramírez F; Section for Chemical Life Sciences, Department of Chemistry, University of Oslo, Oslo NO-0315, Norway.
  • Aizenshtadt A; Department of Analytical Chemistry, Faculty of Science, Charles University, Prague CZ-128 43, Czech Republic.
  • Busek M; Hybrid Technology Hub-Centre of Excellence, Institute of Basic Medical Sciences, Faculty of Medicine, University of Oslo, Oslo 0372, Norway.
  • Krauss SJK; Hybrid Technology Hub-Centre of Excellence, Institute of Basic Medical Sciences, Faculty of Medicine, University of Oslo, Oslo 0372, Norway.
  • Wilson SR; Hybrid Technology Hub-Centre of Excellence, Institute of Basic Medical Sciences, Faculty of Medicine, University of Oslo, Oslo 0372, Norway.
  • Røberg-Larsen H; Hybrid Technology Hub-Centre of Excellence, Institute of Basic Medical Sciences, Faculty of Medicine, University of Oslo, Oslo 0372, Norway.
Anal Chem ; 96(29): 12129-12138, 2024 07 23.
Article em En | MEDLINE | ID: mdl-38985547
ABSTRACT
As organoids and organ-on-chip (OoC) systems move toward preclinical and clinical applications, there is an increased need for method validation. Using a liquid chromatography-mass spectrometry (LC-MS)-based approach, we developed a method for measuring small-molecule drugs and metabolites in the cell medium directly sampled from liver organoids/OoC systems. The LC-MS setup was coupled to an automatic filtration and filter flush system with online solid-phase extraction (SPE), allowing for robust and automated sample cleanup/analysis. For the matrix, rich in, e.g., protein, salts, and amino acids, no preinjection sample preparation steps (protein precipitation, SPE, etc.) were necessary. The approach was demonstrated with tolbutamide and its liver metabolite, 4-hydroxytolbutamide (4HT). The method was validated for analysis of cell media of human stem cell-derived liver organoids cultured in static conditions and on a microfluidic platform according to Food and Drug Administration (FDA) guidelines with regards to selectivity, matrix effects, accuracy, precision, etc. The system allows for hundreds of injections without replacing chromatography hardware. In summary, drug/metabolite analysis of organoids/OoCs can be performed robustly with minimal sample preparation.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Organoides / Fígado Limite: Humans Idioma: En Revista: Anal Chem Ano de publicação: 2024 Tipo de documento: Article País de afiliação: Noruega