Analytical and Clinical Validation of an Automated Microfluidic Interleukin-6 Immunoassay.
Yang, Jianbo; Barbhuiya, Mustafa A; Hamilton, Christopher; Robyak, Kimberly; Zhu, Yusheng.
Afiliação
  • Yang J; Department of Pathology and Laboratory Medicine, Pennsylvania State University College of Medicine, Hershey, PA, USA.
  • Barbhuiya MA; Department of Pathology and Laboratory Medicine, Pennsylvania State University College of Medicine, Hershey, PA, USA.
  • Hamilton C; Department of Pathology and Laboratory Medicine, Pennsylvania State University College of Medicine, Hershey, PA, USA.
  • Robyak K; Department of Pathology and Laboratory Medicine, Pennsylvania State University College of Medicine, Hershey, PA, USA.
  • Zhu Y; Department of Pathology and Laboratory Medicine, Pennsylvania State University College of Medicine, Hershey, PA, USA yzhu@pennstatehealth.psu.edu.
Ann Clin Lab Sci ; 54(3): 388-393, 2024 May.
Article em En | MEDLINE | ID: mdl-39048159
ABSTRACT

OBJECTIVE:

We validated an automated microfluidic interleukin-6 (IL-6) immunoassay on the Ella platform for clinical use.

METHODS:

The assay was validated for precision, lower limit of quantification, analytical measurement range, accuracy, specificity, interference of biotin, tocilizumab, GP130, IL-6Rα, hemolysis, icterus, and lipemia, and establishing the reference value. The clinical performance was evaluated in 96 COVID-19 patients.

RESULTS:

The within-run and between-run coefficient of variations (CV) ranged 1.8%-4.8%. This assay has an analytical measurement range (AMR) 0.7-2652 pg/ml. There was a moderate correlation between Ella IL-6 assay (y) and a Luminex Quantitative Multiplex Bead Assay in a reference lab (x) y=8.561x-475.38, R=0.5047, SEE=1592.8 pg/mL, N=48). Measurement of IL-6 in plasma samples from 45 healthy adults showed the upper limit of reference of 5.0 pg/mL. 95.83% (95% CI 89.67%-98.85%) of COVID-19 patients had IL-6 >5.0 pg/mL. This assay is resistant to the interference from hemoglobin up to 1,144 mg/dL, triglyceride 1,699 mg/dL, bilirubin 35 mg/dL, biotin 1000 ng/mL, tocilizumab 240 µg/mL, IL-11 50,000 pg/mL, GP130 50,000 pg/mL, and IL-6Rα 1,000 pg/mL.

CONCLUSIONS:

The IL-6 assay on the Ella platform is robust with minimum manual operation. The analytical and clinical performance characteristics meet the clinical needs.