Continuous comparison of in vitro and in vivo calibrated transcutaneous oxygen tension with arterial oxygen tension in infants.
Birth Defects Orig Artic Ser ; 15(4): 295-304, 1979.
Article em En | MEDLINE | ID: mdl-534709
Transcutaneous PO2 simultaneously recorded by Drager and Radiometer electrodes on the abdominal skin was compared for six-hour periods with aortic PaO2 recorded by a Searle intravascular oxygen electrode. Ten newborn infants with serious respiratory illnesses, six of whom needed mechanical ventilation, were studied. The skin electrodes were heated to 44 degrees C and calibrated first in vitro, and then in vivo against the infant's PaO2. The results showed that 1) after in vivo calibration both skin electrodes gave an accurate estimate of PaO2 for six hours without resiting of the electrodes; 2) the Radiometer electrode gave as satisfactory an estimate of PaO2 after in vitro as after in vivo calibration; 3) the Drager electrode gave a significantly less accurate estimate of PaO2 after in vitro than after in vivo calibration; 4) no evidence suggesting that peripheral vasoconstriction interfered with the accuracy of estimation of PaO2 by the skin electrodes was found.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxigênio / Doenças Respiratórias / Gasometria / Recém-Nascido de Baixo Peso / Recém-Nascido Prematuro / Doenças do Recém-Nascido Limite: Humans / Newborn Idioma: En Revista: Birth Defects Orig Artic Ser Ano de publicação: 1979 Tipo de documento: Article
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Oxigênio / Doenças Respiratórias / Gasometria / Recém-Nascido de Baixo Peso / Recém-Nascido Prematuro / Doenças do Recém-Nascido Limite: Humans / Newborn Idioma: En Revista: Birth Defects Orig Artic Ser Ano de publicação: 1979 Tipo de documento: Article