Dichromatic absorptiometry of vertebral bone mineral content.
Invest Radiol
; 12(2): 180-4, 1977.
Article
em En
| MEDLINE
| ID: mdl-856759
A dichromatic photon absorptiometric technique for the in vivo measurement of the bone mineral of the spine is discussed. A high purity, high activity 153Gd source which has photons of predominantly 44 and 100 keV was used as the transmission source. The transmission scans were performed on a modified Ohio Nuclear whole body rectillinear scanner. The precision of four measurements on three individuals and three measurements on one individual over a one month period was 2.3% (coefficient of variation). The accuracy determined by measurements of phantoms with known concentrations of dipotassium hydrogen phosphate solutions was 1.2% (standard error of estimate). This technique can be applied in most well equipped nuclear medicine departments.