The calibration of CT Hounsfield units for radiotherapy treatment planning.
Schneider, U; Pedroni, E; Lomax, A.
Afiliação
  • Schneider U; Section of Physics, University of Munich, Garching, Bavaria, Germany.
Phys Med Biol ; 41(1): 111-24, 1996 Jan.
Article em En | MEDLINE | ID: mdl-8685250
Computer tomographic (CT) scans are used to correct for tissue inhomogeneities in radiotherapy treatment planning. In order to guarantee a precise treatment, it is important to obtain the relationship between CT Hounsfield units and electron densities (or proton stopping powers for proton radiotherapy), which is the basic input for radiotherapy planning systems which consider tissue heterogeneities. A method is described to determine improved CT calibrations for biological tissue (a stoichiometric calibration) based on measurements using tissue equivalent materials. The precision of this stoichiometric calibration and the more usual tissue substitute calibration is determined by a comparison of calculated proton radiographic images based on these calibrations and measured radiographs of a biological sample. It has been found that the stoichiometric calibration is more precise than the tissue substitute calibration.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Tomografia Computadorizada por Raios X Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Phys Med Biol Ano de publicação: 1996 Tipo de documento: Article País de afiliação: Alemanha
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Planejamento da Radioterapia Assistida por Computador / Tomografia Computadorizada por Raios X Tipo de estudo: Prognostic_studies Limite: Humans Idioma: En Revista: Phys Med Biol Ano de publicação: 1996 Tipo de documento: Article País de afiliação: Alemanha