Your browser doesn't support javascript.
loading
An implantable centrifugal blood pump with a recirculating purge system (Cool-Seal system).
Artif Organs ; 22(6): 466-74, 1998 Jun.
Article em En | MEDLINE | ID: mdl-9650667
A compact centrifugal blood pump has been developed as an implantable left ventricular assist system. The impeller diameter is 40 mm, and pump dimensions are 55 x 64 mm. This first prototype, fabricated from titanium alloy, resulted in a pump weight of 400 g including a brushless DC motor. The weight of a second prototype pump was reduced to 280 g. The entire blood contacting surface is coated with diamond like carbon (DLC) to improve blood compatibility. Flow rates of over 7 L/min against 100 mm Hg pressure at 2,500 rpm with 9 W total power consumption have been measured. A newly designed mechanical seal with a recirculating purge system (Cool-Seal) is used for the shaft seal. In this seal system, the seal temperature is kept under 40 degrees C to prevent heat denaturation of blood proteins. Purge fluid also cools the pump motor coil and journal bearing. Purge fluid is continuously purified and sterilized by an ultrafiltration unit which is incorporated in the paracorporeal drive console. In vitro experiments with bovine blood demonstrated an acceptably low hemolysis rate (normalized index of hemolysis = 0.005 +/- 0.002 g/100 L). In vivo experiments are currently ongoing using calves. Via left thoracotomy, left ventricular (LV) apex descending aorta bypass was performed utilizing an expanded polytetrafluoroethylene (ePTFE) vascular graft with the pump placed in the left thoracic cavity. In 2 in vivo experiments, the pump flow rate was maintained at 5-9 L/min, and pump power consumption remained stable at 9-10 W. All plasma free Hb levels were measured at less than 15 mg/dl. The seal system has demonstrated good seal capability with negligible purge fluid consumption (<0.5 ml/day). In both calves, the pumps demonstrated trouble free continuous function over 6 month (200 days and 222 days).
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Coração Auxiliar Tipo de estudo: Observational_studies / Prognostic_studies Limite: Animals Idioma: En Revista: Artif Organs Ano de publicação: 1998 Tipo de documento: Article País de afiliação: Japão
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Desenho de Prótese / Coração Auxiliar Tipo de estudo: Observational_studies / Prognostic_studies Limite: Animals Idioma: En Revista: Artif Organs Ano de publicação: 1998 Tipo de documento: Article País de afiliação: Japão