Am J Surg ; 224(1 Pt B): 569-575, 2022 07.
Artículo en Inglés | MEDLINE | ID: mdl-35379482

RESUMEN

BACKGROUND: The COVID-19 pandemic revealed flaws in the stockpiling and distribution of ventilators. In this study, we assessed the durability, sterilizability, and performance of a 3D-printed ventilator. METHODS: SLS-printed devices were dropped from 1.83 m and autoclaved before evaluation on a COVID-19 simulated patient. The respiratory performance of an extrusion-printed device was studied using a variable compliance model. Ranges of sustainable respiratory rates were evaluated as a function of tidal volume. RESULTS: Autoclaving and dropping the device did not negatively impact minute ventilation or PIP for sustained ventilation. Equivalence was significant across all measures except for comparing the autoclaved and dropped with p = 0.06. Extrusion produced ventilators achieved minute ventilation ranging from 4.1 to 12.2 L/min for all simulated compliances; there was an inverse correlation between tidal volume and respiratory rate. CONCLUSION: The CRISIS ventilator is a durable, sterilizable, and reusable 3D-printed ventilator using off-the-shelf materials which could be employed variety of adult lung diseases. Further in-vivo testing is needed.


Asunto(s)
COVID-19 , Pandemias , Adulto , Humanos , Impresión Tridimensional , Frecuencia Respiratoria , Ventiladores Mecánicos