Your browser doesn't support javascript.
loading
Mostrar: 20 | 50 | 100
Resultados 1 - 2 de 2
Filtrar
Más filtros

Banco de datos
Tipo del documento
País de afiliación
Intervalo de año de publicación
1.
Cardiol Young ; 31(11): 1781-1787, 2021 Nov.
Artículo en Inglés | MEDLINE | ID: mdl-33685532

RESUMEN

BACKGROUND: Tetralogy of Fallot is a congenital heart defect diagnosed in infancy. Assessment of right ventricular size and function is important for evaluation of patients with tetralogy of Fallot, but these quantitative measures are challenging by echocardiography. This study evaluates a semi-automated software (EchoInsight®, Epsilon Imaging) by comparing its measures to manual measures in children with tetralogy of Fallot. METHODS: Echocardiographic measurements were performed using manual techniques and semi-automated software. Right ventricular measurements included end-diastolic and end-systolic area, fractional area change, chamber dimensions, and tricuspid annular plane systolic excursion. Reliability, correlation, and agreement between manual and semi-automated measures were assessed. RESULTS: Echocardiograms for 46 patients were analysed. Intra- and inter-observer reliabilities for semi-automated measures were good with intraclass correlation coefficients all over 0.95 and 0.85, respectively. There was high correlation between manual and semi-automated methods for areas and dimensions (r = 0.91-0.98). Tricuspid annular plane systolic excursion measures and fractional area change also correlated, albeit less strongly. The semi-automated measurements of end-systolic and end-diastolic area were a 20 and 47% higher than manual methods, respectively.The semi-automated method yielded a relative 52% lower fractional area change compared to the manual method. CONCLUSIONS: The semi-automated software generates quantitative right ventricular measures in children with tetralogy of Fallot with good reliability and good correlation with manual methods for all measures, but with significant difference between manual and semi-automated techniques for area and functional measures. The specific right ventricular geometry in tetralogy of Fallot children may be why, compared to normal anatomy, greater differences were observed between the two techniques.


Asunto(s)
Tetralogía de Fallot , Disfunción Ventricular Derecha , Niño , Diástole , Ecocardiografía , Ventrículos Cardíacos/diagnóstico por imagen , Humanos , Reproducibilidad de los Resultados , Tetralogía de Fallot/diagnóstico por imagen , Disfunción Ventricular Derecha/diagnóstico por imagen , Función Ventricular Derecha
2.
Cardiol Young ; 29(9): 1149-1159, 2019 Sep.
Artículo en Inglés | MEDLINE | ID: mdl-31412970

RESUMEN

BACKGROUND: Assessment of right ventricular size and function is an important part of the clinical cardiac evaluation; however, these quantitative measures are challenging by echocardiography. Automated software could be useful in place of manual measurements and qualitative assessment. This study evaluates a semi-automated software by comparing automated to manual measures in normal children. METHODS: Neonates to adolescents with normal echocardiograms were prospectively enrolled. Measurements were performed using manual techniques and semi-automated software (EchoInsight®, Epsilon Imaging, Ann Arbor, Michigan, United States of America). Right ventricular measurements included end-diastolic and end-systolic area, fractional area change, chamber dimensions, and tricuspid annular plane systolic excursion. Agreement between manual and semi-automated measures was compared. RESULTS: Echocardiograms for 233 patients were included in the analysis. Intra- and inter-observer reliabilities for semi-automated measures were good with intraclass correlation coefficients all over 0.9 and 0.85, respectively. There was very strong correlation between manual and semi-automated methods for areas and dimensions (r = 0.93-0.99) and low bias (1.4-10.8%). For functional measures, tricuspid annular plane systolic excursion measures correlated well (r = 0.84), but fractional area change did not (r = 0.50). Both demonstrated significant bias (33.5-43.0%). The semi-automated method consistently underestimated fractional area change with a mean of 26.6% versus a manual mean of 36.1%. CONCLUSIONS: The semi-automated software is capable of generating quantitative right ventricular measures in children with good reliability. The software demonstrates very good correlation and low bias when compared to manual methods for right ventricular areas and dimensions. There is a significant difference between manual and semi-automated techniques for the functional measures.


Asunto(s)
Ecocardiografía/métodos , Ventrículos Cardíacos/diagnóstico por imagen , Aumento de la Imagen/métodos , Función Ventricular Derecha/fisiología , Adolescente , Niño , Preescolar , Femenino , Estudios de Seguimiento , Humanos , Lactante , Recién Nacido , Masculino , Estudios Prospectivos , Valores de Referencia , Reproducibilidad de los Resultados
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA