1.
J Radiol Prot ; 41(2)2021 06 01.
Artículo en Inglés | MEDLINE | ID: mdl-33873158

RESUMEN

There is currently general concern over low-level radioactive wastewater from the development of nuclear industry. In this paper, a method based on an ultralow-level liquid scintillation spectrometer for measuring uranium radioactivity in low-level radioactive wastewater is proposed. This method can easily and quickly measure the radioactivity level of uranium in samples and can even distinguish the main isotopes of uranium. The liquid scintillation method directly provides results in units of radioactivity activity concentration, which are more convenient for comparison with relevant national standards to determine whether the emission standards are met. The lowest limit of detection of this method is 0.014 Bq l-1within 600 min.


Asunto(s)
Monitoreo de Radiación , Radiactividad , Contaminantes Radiactivos del Agua , Conteo por Cintilación , Aguas Residuales , Contaminantes Radiactivos del Agua/análisis
SELECCIÓN DE REFERENCIAS
DETALLE DE LA BÚSQUEDA