Magn Reson Imaging ; 105: 10-16, 2024 Jan.
Artículo en Inglés | MEDLINE | ID: mdl-37863374

RESUMEN

Diffusion-weighted double-echo steady-state (dwDESS) MRI with bipolar diffusion gradients is a promising candidate to obtain diffusion weighted images (DWI) free of geometric distortions and with low motion sensitivity. However, a wider clinical application of dwDESS is currently hindered as no method is reported to explicitly calculate the effective b-value of the obtained DWI from the diffusion-gradients applied in the sequence. To this end, a previously described signal model was adapted for dwDESS with bipolar diffusion gradients, which allows to estimate an effective b-value, dubbed b'. Evaluation in phantom examinations was performed on a clinical 1.5 T MR system. Experimental results were compared with theoretical predictions, including the apparent diffusion coefficient (ADC) based on b-values from a standard EPI-DWI sequence and ADC' based on the effective b' from the dwDESS sequence. The adapted signal model was able to describe the experimental results, and the obtained values of ADC' were in line with conventional ADC measurements.


Asunto(s)
Imagen de Difusión por Resonancia Magnética , Imagen por Resonancia Magnética , Imagen por Resonancia Magnética/métodos , Imagen de Difusión por Resonancia Magnética/métodos , Movimiento (Física) , Fantasmas de Imagen , Imagen Eco-Planar