RESUMO
The CRESST experiment is designed to search for weakly interacting massive particle dark matter with cryogenic detectors. CRESST II will use up to 33 CaWO(4) crystals with a total mass of approximately 10 kg. These many detectors require a readout system based on 66-channel superconducting quantum interference devices (SQUIDs). In this article we report on the development of a modular superconducting connector for the 66-channel SQUID readout circuit. We show that the technique developed reliably produces superconducting contacts.
Assuntos
Eletroquímica/instrumentação , Eletrodos , Eletrônica/instrumentação , Transdutores , Condutividade Elétrica , Eletroquímica/métodos , Desenho de Equipamento , Análise de Falha de Equipamento , Teoria Quântica , Reprodutibilidade dos Testes , Sensibilidade e EspecificidadeRESUMO
One of the most intriguing open questions in modern particle physics is the nature of the dark matter in our universe. As hypothetical weakly interacting massive particles (WIMPs) do interact with ordinary matter extremely rarely, their observation requires a very low-background detector environment regarding radioactivity as well as an advanced detector technique that allows for active discrimination of the still present radioactive contaminations. The CRESST experiment uses detectors operating at milli-Kelvin temperature. Energy deposition in the detectors is recorded via the simultaneous measurement of a phonon-mediated signal and scintillation emitted by the CaWO(4) crystal targets. The entire setup is made of carefully selected materials. In this note we report on the development of ultra-pure bronze (CuSn(6)) wire in small quantities for springs and clamps that are currently being used in the CRESST II setup.