Improved sensitivity in detection of chlormequat by liquid chromatography-mass spectrometry.
Evans, C S; Startin, J R; Goodall, D M; Keely, B J.
Afiliación
  • Evans CS; Chemistry Department, University of York, Heslington, UK.
J Chromatogr A ; 897(1-2): 399-404, 2000 Nov 03.
Article en En | MEDLINE | ID: mdl-11128224
Two published separations, using electrospray mass spectrometry (ES-MS), exhibit significant differences in limits of detection (LODs) for chlormequat cation in pear. Separation on ODS1, confirmed to result from ion-exchange, gives shorter analysis times and calibration over a wider concentration range than on an SCX cation-exchange column. The superior LOD using ODS1 (0.04 ng ml(-1) vs. 1.0 ng ml(-1)) results mainly from better chromatographic peak shape. Separation on ODS1 combined with optimised ES-MS detection allows direct quantification of chlormequat on an ion trap instrument at levels lower than those required for residue analysis in foods and also in drinking water.