Dosimetric characterization of round HDR 192Ir accuboost applicators for breast brachytherapy.
Med Phys
; 36(11): 5027-32, 2009 Nov.
Article
en En
| MEDLINE
| ID: mdl-19994513
PURPOSE: The AccuBoost brachytherapy system applies HDR 192Ir beams peripherally to the breast using collimating applicators. The purpose of this study was to benchmark Monte Carlo simulations of the HDR 192Ir source, to dosimetrically characterize the round applicators using established Monte Carlo simulation and radiation measurement techniques and to gather data for clinical use. METHODS: Dosimetric measurements were performed in a polystyrene phantom, while simulations estimated dose in air, liquid water, polystyrene and ICRU 44 breast tissue. Dose distribution characterization of the 4-8 cm diameter collimators was performed using radiochromic EBT film and air ionization chambers. RESULTS: The central axis dose falloff was steeper for the 4 cm diameter applicator in comparison to the 8 cm diameter applicator, with surface to 3 cm depth-dose ratios of 3.65 and 2.44, respectively. These ratios did not considerably change when varying the phantom composition from breast tissue to polystyrene, phantom thickness from 4 to 8 cm, or phantom radius from 8 to 15 cm. Dose distributions on the central axis were fitted to sixth-order polynomials for clinical use in a hand calculation spreadsheet (i.e., nomogram). Dose uniformity within the useful applicator apertures decreased as depth-dose increased. CONCLUSIONS: Monte Carlo benchmarking simulations of the HDR 192Ir source using the MCNP5 radiation transport code indicated agreement within 1% of the published results over the radial/angular region of interest. Changes in phantom size and radius did not cause noteworthy changes in the central axis depth-dose. Polynomial fit depth-dose curves provide a simple and accurate basis for a nomogram.
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Radiometría
/
Planificación de la Radioterapia Asistida por Computador
/
Braquiterapia
/
Neoplasias de la Mama
/
Radioisótopos de Iridio
Tipo de estudio:
Health_economic_evaluation
Límite:
Female
/
Humans
Idioma:
En
Revista:
Med Phys
Año:
2009
Tipo del documento:
Article
País de afiliación:
Estados Unidos