Diffusion-weighted imaging of the abdomen at 3 T: image quality comparison with 1.5-T magnet using 3 different imaging sequences.
Saremi, Farhood; Jalili, Mehdi; Sefidbakht, Sepideh; Channual, Stephanie; Quane, Lisa; Naderi, Nassim; Schultze-Haakh, Helmuth; Torrone, Maria.
Afiliación
  • Saremi F; Department of Radiological Sciences of University of California, Irvine, CA 92868-3298, USA. fsaremi@uci.edu
J Comput Assist Tomogr ; 35(3): 317-25, 2011.
Article en En | MEDLINE | ID: mdl-21586923
ABSTRACT

OBJECTIVE:

This study aimed to perform comparisons between diffusion-weighted imaging (DWI) sequences at 3 T with 1.5 T.

METHODS:

Thirteen healthy volunteers underwent abdominal DWI on both 3- and 1.5-T magnets using 3 sequences including breath hold without parallel imaging (PI), breath hold with PI, and free breathing with PI at b50 and b1000. Artifacts and subjective image quality scores, signal intensity, and apparent diffusion coefficient were compared.

RESULTS:

For breath hold without PI, higher artifact was noted at 3 T b50 compared with 1.5 T (P < 0.0001). For b50 and b1000 breath hold with PI, artifacts were not different between the magnets, but image quality was better at 3 T (P = 0.04 and P = 0.02, respectively). For b50 and b1000 free breathing sequences, artifact and image quality scores were significantly better at 1.5 T. For breath hold acquisitions, the signal-to-noise ratio of gallbladder, kidneys, and pancreas was generally higher and that of the liver was lower on 3 T. Imaging at 3 T showed significantly higher image quality and lower artifacts for breath hold with PI compared with free breathing. Most apparent diffusion coefficients were not significantly different between the 2 magnets (P > 0.05).

CONCLUSIONS:

Three-tesla magnets can provide good images using breath hold with PI sequence.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Imagen de Difusión por Resonancia Magnética / Abdomen Tipo de estudio: Observational_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Comput Assist Tomogr Año: 2011 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Imagen de Difusión por Resonancia Magnética / Abdomen Tipo de estudio: Observational_studies Límite: Adult / Aged / Female / Humans / Male / Middle aged Idioma: En Revista: J Comput Assist Tomogr Año: 2011 Tipo del documento: Article País de afiliación: Estados Unidos