Your browser doesn't support javascript.
loading
Comparison of outcomes in patients with abandoned versus extracted implantable cardioverter defibrillator leads.
Amelot, Mathieu; Foucault, Anthony; Scanu, Patrice; Gomes, Sophie; Champ-Rigot, Laure; Pellissier, Arnaud; Milliez, Paul.
Afiliación
  • Amelot M; Cardiology Department, Côte-de-Nacre University Hospital, avenue de la Côte-de-Nacre, 14033 Caen, France.
Arch Cardiovasc Dis ; 104(11): 572-7, 2011 Nov.
Article en En | MEDLINE | ID: mdl-22117909
ABSTRACT

BACKGROUND:

Despite the increased number of implantable cardioverter defibrillator (ICD) recipients and the frequent need for device upgrading and/or occurrence of lead malfunction, the optimal approach to managing abandoned leads remains debated.

AIMS:

To determine the rate and type of complications related to either abandoned or extracted ICD leads.

METHODS:

Patients with abandoned or extracted leads were identified retrospectively. Patient medical records were reviewed to assess long-term lead or device malfunction, defibrillation test values before and after lead abandonment or extraction, and appropriateness of delivered shocks and subsequent surgical procedures related to devices or leads.

RESULTS:

A total of 58 ICD patients with 47 extracted and 34 abandoned leads were identified. After a mean follow-up of 3.2 ± 2.6 years, the defibrillation test was not affected by either abandoned or extracted leads (23.4 ± 6.6 J vs 25.4 ± 4.9 J, respectively; P = 0.24). There were no differences in the number of ICD-related surgical procedures after extracting versus abandoning leads (22% vs 12%, respectively; P = 0.3) or in the thromboembolic event rate (7.7% vs 6.3%; P = 0.83). During follow-up, no differences in the occurrence of major complications or appropriate/inappropriate shocks were observed between patients with or without abandoned leads.

CONCLUSION:

We observed no difference in rates of immediate or medium-term complications between extracting versus abandoning leads. Lead abandonment remains an alternative and safe option when extraction does not appear mandatory according to the age of the leads or experience of the operating centre.
Asunto(s)

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Falla de Prótesis / Cardioversión Eléctrica / Desfibriladores Implantables / Remoción de Dispositivos / Insuficiencia Cardíaca Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Aged / Humans / Middle aged País/Región como asunto: Europa Idioma: En Revista: Arch Cardiovasc Dis Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2011 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Falla de Prótesis / Cardioversión Eléctrica / Desfibriladores Implantables / Remoción de Dispositivos / Insuficiencia Cardíaca Tipo de estudio: Etiology_studies / Observational_studies / Prognostic_studies / Risk_factors_studies Límite: Adult / Aged / Humans / Middle aged País/Región como asunto: Europa Idioma: En Revista: Arch Cardiovasc Dis Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA Año: 2011 Tipo del documento: Article País de afiliación: Francia