Your browser doesn't support javascript.
loading
Magnetization transfer from inhomogeneously broadened lines (ihMT): Experimental optimization of saturation parameters for human brain imaging at 1.5 Tesla.
Girard, Olivier M; Prevost, Valentin H; Varma, Gopal; Cozzone, Patrick J; Alsop, David C; Duhamel, Guillaume.
Afiliación
  • Girard OM; Aix Marseille Université, CRMBM-CEMEREM UMR 7339, CNRS - AMU, Marseille, France.
  • Prevost VH; Aix Marseille Université, CRMBM-CEMEREM UMR 7339, CNRS - AMU, Marseille, France.
  • Varma G; Department of Radiology, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Massachusetts, USA.
  • Cozzone PJ; Aix Marseille Université, CRMBM-CEMEREM UMR 7339, CNRS - AMU, Marseille, France.
  • Alsop DC; Department of Radiology, Beth Israel Deaconess Medical Center and Harvard Medical School, Boston, Massachusetts, USA.
  • Duhamel G; Aix Marseille Université, CRMBM-CEMEREM UMR 7339, CNRS - AMU, Marseille, France.
Magn Reson Med ; 73(6): 2111-21, 2015 Jun.
Article en En | MEDLINE | ID: mdl-24962257
ABSTRACT

PURPOSE:

Recently a new MR endogenous contrast mechanism was reported. It allows specifically imaging the magnetization transfer (MT) effect arising from inhomogeneously broadened components of the NMR spectrum, and was hence dubbed ihMT. Such unique NMR lineshape properties are presumably occurring in myelin because of its specifically ordered, multilayered sheath structure. Here, optimization of a pulsed ihMT preparation module is presented to provide guidance for future studies and improve the understanding of underlying contrast mechanisms.

METHODS:

This study was performed at 1.5 Tesla on healthy volunteers. A pulsed ihMT preparation was implemented in combination with a HASTE readout module. The pulse width, interpulse repetition time, total saturation duration and RF saturation power were considered for optimization of the ihMT sensitivity and contrast.

RESULTS:

An optimal configuration of the preparation module was derived, leading to 10% ihMT signal in internal capsule (relative to unsaturated data) and around 200% signal increase relative to gray matter, i.e., approximately 10-fold superior contrast compared with conventional MT ratios, measured under similar experimental conditions.

CONCLUSION:

Overall the ihMT sequence was robust, sensitive and very specific for white matter. These findings suggest great potential for assessing brain myelination and for better characterization of myelin related disorders.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Mapeo Encefálico / Imagen por Resonancia Magnética / Aumento de la Imagen / Sustancia Blanca Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2015 Tipo del documento: Article País de afiliación: Francia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Mapeo Encefálico / Imagen por Resonancia Magnética / Aumento de la Imagen / Sustancia Blanca Tipo de estudio: Diagnostic_studies Límite: Humans Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2015 Tipo del documento: Article País de afiliación: Francia