Automated antinuclear immunofluorescence antibody analysis is a reliable approach in routine clinical laboratories.
Zheng, Bing; Li, Enling; Zhu, Haoming; Lu, Jingbo; Shi, Xinming; Zhang, Jie; Li, Min.
Afiliación
  • Zheng B; .
  • Li E; .
  • Zhu H; .
  • Lu J; .
  • Shi X; .
  • Zhang J; .
  • Li M; .
Clin Chem Lab Med ; 55(12): 1922-1930, 2017 Oct 26.
Article en En | MEDLINE | ID: mdl-28467309
BACKGROUND: Indirect immunofluorescence (IIF) assays are recommended as the gold standard method for the detection of antinuclear antibodies (ANAs). This study aimed to investigate the reliability of an automated system. METHODS: We compared 3745 serum samples using NOVA View archived images with manual analysis via microscopy. A custom cutoff value was established to distinguish ANA titers and was validated in two clinical laboratories. The automatic ANA pattern recognition system was evaluated, and all ANA-positive sera were subjected to two commercial ANA IIF kits to compare the consistency of the pattern interpretation results. For inconsistent patterns, a third ANA IIF testing kit was utilized. RESULTS: Agreement of the interpretation of the ANA IIF test using the platform of NOVA View and manual microscopy was 96.9%. The local cutoff value to discriminate ANA titers in four main ANA patterns was calculated based on 1390 serum samples. In our laboratory, the titer prediction accuracy was superior to the preset cutoff in NOVA View (p<0.01); the performance was similar in another laboratory (p=0.11). The automatic pattern recognition accuracies of speckled, homogeneous, centromere, nucleolar and nuclear dot patterns were 62.7%, 57.4%, 92.6%, 30.5% and 27.3%, respectively. The consistency of the pattern interpretation results between INOVA and MBL kits was 95.3%. CONCLUSIONS: It is necessary to establish a custom value-added ANA report. However, confirmation of the digital immunofluorescence images by expert technicians was essential, and suspect results of an ANA pattern should be reconfirmed by another commercial ANA IIF kit to achieve more reliable results.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Automatización / Anticuerpos Antinucleares / Enfermedades del Tejido Conjuntivo / Técnica del Anticuerpo Fluorescente Indirecta / Técnicas de Laboratorio Clínico / Pruebas Diagnósticas de Rutina Tipo de estudio: Diagnostic_studies / Guideline / Prognostic_studies Límite: Humans Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2017 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Automatización / Anticuerpos Antinucleares / Enfermedades del Tejido Conjuntivo / Técnica del Anticuerpo Fluorescente Indirecta / Técnicas de Laboratorio Clínico / Pruebas Diagnósticas de Rutina Tipo de estudio: Diagnostic_studies / Guideline / Prognostic_studies Límite: Humans Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2017 Tipo del documento: Article