Radiation dose and image quality with new protocol in lower extremity computed tomography angiography.
Schicchi, Nicolò; Fogante, Marco; Oliva, Matteo; Esposto Pirani, Paolo; Agliata, Giacomo; Giuseppetti, Gian Marco; Giovagnoni, Andrea.
Afiliación
  • Schicchi N; Department of Radiology, Azienda Ospedaliero-Universitaria, Ospedali Riuniti di Ancona, Via Conca 71, 60126, Torrette, Ancona, Italy.
  • Fogante M; Department of Radiology-School of Radiology, Università Politecnica delle Marche, Ancona, Italy. marco.fogante89@gmail.com.
  • Oliva M; Department of Radiology-School of Radiology, Università Politecnica delle Marche, Ancona, Italy.
  • Esposto Pirani P; Department of Radiology, Azienda Ospedaliero-Universitaria, Ospedali Riuniti di Ancona, Via Conca 71, 60126, Torrette, Ancona, Italy.
  • Agliata G; Department of Radiology, Azienda Ospedaliero-Universitaria, Ospedali Riuniti di Ancona, Via Conca 71, 60126, Torrette, Ancona, Italy.
  • Giuseppetti GM; Department of Radiology, Azienda Ospedaliero-Universitaria, Ospedali Riuniti di Ancona, Via Conca 71, 60126, Torrette, Ancona, Italy.
  • Giovagnoni A; Department of Radiology, Azienda Ospedaliero-Universitaria, Ospedali Riuniti di Ancona, Via Conca 71, 60126, Torrette, Ancona, Italy.
Radiol Med ; 124(3): 184-190, 2019 Mar.
Article en En | MEDLINE | ID: mdl-30478814
ABSTRACT

PURPOSE:

To compare radiation dose and image quality of lower extremity computed tomography angiography (CTA) between cranio-caudal acquisition with single-source CT (SSCT) and flash caudo-cranial acquisition with dual-source CT (DSCT). MATERIALS AND

METHODS:

In this prospective study, 60 patients were randomly assigned to Group A (control) or Group B (experimental) to undergo lower extremity CTA for peripheral obliterative arterial disease. Group A received protocol 1 (P1) with SSCT cranio-caudal acquisition. Group B received protocol (P2) with DSCT flash caudo-cranial acquisition. Intravascular attenuation (IVA), signal-to-noise ratio (SNR), contrast-to-noise ratio (CNR) and image noise were compared. Two radiologists assessed the image quality. Computed tomography volume dose index (CTDIvol) and dose-length product (DLP) were also compared.

RESULTS:

IVA with P2 was higher than with P1 (586.8 ± 140.3 vs. 496.1 ± 129.3 HU, p = 0.011), as was SNR (33.0 ± 11.3 vs. 27.4 ± 12.3; p = 0.042), CNR (30.1 ± 13.3 vs. 24.2 ± 10.3; p = 0.029) and image quality score of small arteries below the knee (3.8 ± 0.2 vs. 3.1 ± 0.2; p = 0.001). Radiation dose was significantly lower in P2 than in P1 with CTDIvol reduction of 40.9% (1.3 ± 0.1 vs. 2.2 ± 0.3 mGy; p = 0.006) and DLP reduction of 42.8% (148.7 ± 21.9 vs 260.2 ± 59.1 mGy * cm; p = 0.018).

CONCLUSION:

Lower extremity CTA with DSCT flash caudo-cranial acquisition allows lower radiation dose with higher IVA, SNR, CNR and better image quality for small arteries below the knee than SSCT cranio-caudal acquisition.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Dosis de Radiación / Extremidad Inferior / Enfermedad Arterial Periférica / Angiografía por Tomografía Computarizada Tipo de estudio: Clinical_trials / Guideline / Observational_studies Límite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Radiol Med Año: 2019 Tipo del documento: Article País de afiliación: Italia

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Dosis de Radiación / Extremidad Inferior / Enfermedad Arterial Periférica / Angiografía por Tomografía Computarizada Tipo de estudio: Clinical_trials / Guideline / Observational_studies Límite: Adolescent / Adult / Aged / Aged80 / Female / Humans / Male / Middle aged Idioma: En Revista: Radiol Med Año: 2019 Tipo del documento: Article País de afiliación: Italia