Preliminary results of the CeraflexTM PDA occluder and device behaviour during releasing.
Epçaçan, Serdar; Bulut, Mustafa Orhan; Yücel, Ilker Kemal; Ceviz, Naci; Çelebi, Ahmet.
Afiliación
  • Epçaçan S; Department of Pediatric Cardiology,University of Health Sciences,Van Training and Research Hospital, Van,Turkey.
  • Bulut MO; Department of Pediatric Cardiology,University of Health Sciences,Dr. Siyami Ersek Thoracic and Cardiovascular Surgery Training and Research Hospital, Istanbul,Turkey.
  • Yücel IK; Department of Pediatric Cardiology,University of Health Sciences,Dr. Siyami Ersek Thoracic and Cardiovascular Surgery Training and Research Hospital, Istanbul,Turkey.
  • Ceviz N; Medical Faculty, Department of Pediatric Cardiology,Ataturk University,Erzurum,Turkey.
  • Çelebi A; Department of Pediatric Cardiology,University of Health Sciences,Dr. Siyami Ersek Thoracic and Cardiovascular Surgery Training and Research Hospital, Istanbul,Turkey.
Cardiol Young ; 29(4): 519-523, 2019 04.
Article en En | MEDLINE | ID: mdl-31032755
ABSTRACT

INTRODUCTION:

The CeraFlexTM PDA occluder is a new flexible device with a unique delivery system that may be beneficial with regard to not changing the device position after releasing. We prospectively evaluate the efficacy of the device and also the device behaviour patterns during release.

METHODS:

The study included 21 patients. Their median age was 1.2 years (from 6 months to 28 years) and weight was 9.6 kg (from 5.4 to 82 kg). All of the ducts were conical except one atypical ductus. Median ductal diameter at the pulmonary end was 3.8 mm (from 2.2 to 8.2 mm). The ductus was closed using an antegrade approach, but special attention was paid to the patterns of device behaviour during and just after releasing.

RESULTS:

Three different modes of device behaviour were observed during and just after releasing (1) Neither difficulty nor change of position in 13 patients (62%), (2) a little difficulty in releasing but no change of position in 6 (29%), and (3) change of the device position in 2 (9%). There was no residual shunt on the next day except in one patient, in whom late device embolisation occurred. The device was retrieved and another, bigger device implanted.

CONCLUSION:

The CeraFlexTM PDA occlude device seems to be safe and efficacious for patent ductus arteriosus closure. Its unique delivery system generally fixes the device in a stable position that does not change after release (91%). Minor difficulty in releasing is not uncommon; however, the major disadvantage is the need for larger sheaths for delivery.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Cateterismo Cardíaco / Conducto Arterioso Permeable / Dispositivo Oclusor Septal Límite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Infant / Male País/Región como asunto: Asia Idioma: En Revista: Cardiol Young Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / PEDIATRIA Año: 2019 Tipo del documento: Article País de afiliación: Turquía

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Cateterismo Cardíaco / Conducto Arterioso Permeable / Dispositivo Oclusor Septal Límite: Adolescent / Adult / Child / Child, preschool / Female / Humans / Infant / Male País/Región como asunto: Asia Idioma: En Revista: Cardiol Young Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / PEDIATRIA Año: 2019 Tipo del documento: Article País de afiliación: Turquía