The established and the challenger: A direct comparison of current cryoballoon technologies for pulmonary vein isolation.
Moser, Fabian; Rottner, Laura; Moser, Julia; Schleberger, Ruben; Lemoine, Marc; Münkler, Paula; Dinshaw, Leon; Kirchhof, Paulus; Reissmann, Bruno; Ouyang, Feifan; Rillig, Andreas; Metzner, Andreas.
Afiliación
  • Moser F; Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
  • Rottner L; Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
  • Moser J; Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
  • Schleberger R; Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
  • Lemoine M; Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
  • Münkler P; Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
  • Dinshaw L; Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
  • Kirchhof P; Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
  • Reissmann B; Institute of Cardiovascular Sciences, College of Medical and Dental Sciences, University of Birmingham, Birmingham, UK.
  • Ouyang F; Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
  • Rillig A; Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
  • Metzner A; Department of Cardiology, University Heart and Vascular Center, Hamburg, Germany.
J Cardiovasc Electrophysiol ; 33(1): 48-54, 2022 01.
Article en En | MEDLINE | ID: mdl-34766404
ABSTRACT

INTRODUCTION:

Cryoballoon (CB) ablation for pulmonary vein isolation (PVI) is an effective treatment of atrial fibrillation (AF). Recently, a novel cryoablation system was introduced. The aim of the study was to compare the safety, efficacy and biophysical characteristics of a novel cryoablation system (POLARx™; Boston Scientific) to a commonly used and clinically well characterized system (Arctic Front Advance Pro™, AFA; Medtronic). METHODS AND

RESULTS:

Fifty consecutive patients with symptomatic AF, who underwent CB-based ablation with the POLARx were compared to 50 consecutive patients treated with the AFA. Acute PVI was achieved in 99.8% (POLARx 99.5%, AFA 100%, p = 1.00). Time to isolation (TTI) was comparable in both groups (POLARx 35 [27, 48] s, AFA 30 [21, 43] s, p = 0.165). The POLARx showed a lower balloon temperature at TTI (POLARx -44 [-50, -36] °C, AFA -31 [-38, -21] °C, p < 0.001) and lower nadir temperature (POLARx -60 [-65, -55] °C, AFA -48 [-54, -45] °C, p < 0.001). Procedure time (POLARx 80 [60, 105] min, AFA 62 [42, 80] min, p < 0.001), fluoroscopy time (POLARx 17 [13, 22] min, AFA 11 [7, 16] min, p < 0.001) and freeze cycles per patient (POLARx 5 [4, 6], AFA 4.5 [4, 5], p = 0.002) were higher in the POLARx group. Two cerebral ischemic events occurred in the POLARx group, two patients in each group had phrenic nerve injury.

CONCLUSION:

Both systems enable effective isolation of pulmonary veins. The POLARx required longer procedure and fluoroscopy times. Larger, prospective and randomized studies are needed to assess long-term efficacy and safety of this technology.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Venas Pulmonares / Fibrilación Atrial / Ablación por Catéter / Criocirugía Tipo de estudio: Clinical_trials / Observational_studies Límite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Venas Pulmonares / Fibrilación Atrial / Ablación por Catéter / Criocirugía Tipo de estudio: Clinical_trials / Observational_studies Límite: Humans Idioma: En Revista: J Cardiovasc Electrophysiol Asunto de la revista: ANGIOLOGIA / CARDIOLOGIA / FISIOLOGIA Año: 2022 Tipo del documento: Article País de afiliación: Alemania