Comparison of noninvasive continuous arterial blood pressure measured by NICAP with arterial line in elderly patients.
BMC Geriatr
; 22(1): 108, 2022 02 07.
Article
en En
| MEDLINE
| ID: mdl-35130866
ABSTRACT
BACKGROUND:
Non-Invasive Continuous Arterial Pressure system (NICAP) allows continuous monitoring, timely detection of hypotension, and avoiding risks from invasive procedures. A previous study showed good comparability of NICAP with arterial line in people with no evidence of cardiovascular disease. Therefore, the goal of this study was to investigate whether NICAP could be accurately applied to elderly patients.METHODS:
In this single-centered observational study, forty-one patients above 65 undergoing elective surgeries requiring artery catheterizations were enrolled from July 17, 2020, to June 25, 2021. Radial artery cannulation and NICAP monitoring were started before anesthesia. Blood pressure during the anesthesia induction and the whole surgery, trend of blood pressure changes, time needed for establishing continuous monitoring, and complications were recorded.RESULTS:
A total of 6751 valid pairs of blood pressure measurements were analyzed. In the Bland-Altman analysis, the arithmetic means for systolic, diastolic, and mean arterial pressure were 2.2, 3.3, and 2.8 mmHg, respectively. NICAP and arterial line correlation coefficients for systolic, diastolic, and mean arterial pressure were 0.49, 0.33, and 0.45, respectively. In the trending analysis, the polar concordance rates at 30 degrees were 70.9% for systolic, 67.7% for diastolic, and 69.3% for mean arterial blood pressure. During the anesthesia induction, the arithmetic means for systolic, diastolic, and mean arterial pressure in the Bland-Altman analysis were 1.7, -0.2, and 0.5 mmHg, respectively. NICAP and arterial line correlation coefficients for systolic, diastolic, and mean arterial pressure were 0.78, 0.61 and 0.75, respectively. No severe complications occurred.CONCLUSIONS:
NICAP has a poor correlation with the arterial line in elderly patients for the whole surgery or during anesthesia induction. Moreover, it showed poor comparability in the detection of blood pressure change trends with arterial lines. Our findings suggest that NICAP might not be sufficiently accurate to be applied clinically in elderly patients with comorbidities. More accurate calibration and iteration are needed.Palabras clave
Texto completo:
1
Colección:
01-internacional
Banco de datos:
MEDLINE
Asunto principal:
Dispositivos de Acceso Vascular
/
Presión Arterial
Tipo de estudio:
Observational_studies
Límite:
Aged
/
Humans
Idioma:
En
Revista:
BMC Geriatr
Asunto de la revista:
GERIATRIA
Año:
2022
Tipo del documento:
Article
País de afiliación:
China