A comparison of four liquid chromatography-mass spectrometry platforms for the analysis of zeranols in urine.
Lazofsky, Abigail; Brinker, Anita; Rivera-Núñez, Zorimar; Buckley, Brian.
Afiliación
  • Lazofsky A; Environmental and Occupational Health Sciences Institute, Rutgers University, Piscataway, NJ, 08854, USA.
  • Brinker A; Environmental and Occupational Health Sciences Institute, Rutgers University, Piscataway, NJ, 08854, USA.
  • Rivera-Núñez Z; Environmental and Occupational Health Sciences Institute, Rutgers University, Piscataway, NJ, 08854, USA.
  • Buckley B; Department of Biostatistics and Epidemiology, Rutgers School of Public Health, Rutgers University, Piscataway, NJ, 08854, USA.
Anal Bioanal Chem ; 415(20): 4885-4899, 2023 Aug.
Article en En | MEDLINE | ID: mdl-37432442
ABSTRACT
Targeted biomonitoring studies quantifying the concentration of zeranols in biological matrices have focused on liquid chromatography interfaced to mass spectrometry (LC-MS). The MS platform for measurement, quadrupole, time-of-flight (ToF), ion trap, etc., is often chosen based on either sensitivity or selectivity. An instrument performance comparison of the benefits and limitations using matrix-matched standards containing 6 zeranols on 4 MS instruments, 2 low-resolution (linear ion traps), and 2 high-resolution (Orbitrap and ToF) was undertaken to identify the best measurement platform for multiple biomonitoring projects characterizing the endocrine disruptive properties of zeranols. Analytical figures of merit were calculated for each analyte to compare instrument performance across platforms. The calibration curves had correlation coefficients r = 0.989 ± 0.012 for all analytes and LODs and LOQs were ranked for sensitivity Orbitrap > LTQ > LTQXL > G1 (V mode) > G1 (W mode). The Orbitrap had the smallest measured variation (lowest %CV), while the G1 had the highest. Instrumental selectivity was calculated using full width at half maximum (FWHM) and as expected, the low-resolution instruments had the broadest spectrometric peaks, concealing coeluting peaks under the same mass window as the analyte. Multiple peaks from concomitant ions, unresolved at low resolution (within a unit mass window), were present but did not match the exact mass predicted for the analyte. For example, the high-resolution platforms were able to differentiate between a concomitant peak at 319.1915 from the analyte at 319.1551, included in low-resolution quantitative analyses demonstrating the need to consider coeluting interfering ions in biomonitoring studies. Finally, a validated method using the Orbitrap was applied to human urine samples from a pilot cohort study.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Zeranol Tipo de estudio: Observational_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Anal Bioanal Chem Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Zeranol Tipo de estudio: Observational_studies / Risk_factors_studies Límite: Humans Idioma: En Revista: Anal Bioanal Chem Año: 2023 Tipo del documento: Article País de afiliación: Estados Unidos