Phantom study investigating the repeatability of radiomic features with alteration of image acquisition parameters in magnetic resonance imaging.
J Med Imaging Radiat Sci
; 55(1): 19-28, 2024 Mar.
Article
en En
| MEDLINE
| ID: mdl-37932212
ABSTRACT
BACKGROUND:
Magnetic resonance imaging (MRI) has many different alterable parameters that affect how an image appears. This is relevant in radiomics which produces quantitative features through analysis of medical images. One significant acknowledged limitation of radiomics is repeatability. This phantom study aims to further investigate the repeatability of radiomic features (RaF), within MRI, across a range of different echo (TE) and repetition times (TR).METHODS:
A phantom was scanned 10 times under identical conditions on a 3T scanner using head coil over 4 months. The TE ranged from 80 to 110 ms while the TR from 3000 to 5000 ms. Radiomics analysis was performed on the same segmented section of the phantom across all TE and TR combinations. Intraclass Correlation Coefficient (ICC) was calculated across the different TE and TR ranges to investigate the repeatability of RaF.RESULTS:
Of 1596 features calculated, 187 features had ICC >0.9 across the range of TE, while 82 features had an ICC >0.9 across a range of TR. 664 had ICC >0.75 across the range of TEs, with 541 across the range of TR values. There was an overlap of 51 features with ICC >0.9.CONCLUSION:
Repeatability of RaF in MRI is dependent on imaging parameters and careful consideration of these, in combination with variable selection, is required when applying radiomics to MRI.Palabras clave