Targeted ultra performance liquid chromatography tandem mass spectrometry procedures for the diagnosis of inborn errors of metabolism: validation through ERNDIM external quality assessment schemes.
Oliva, Clara; Arias, Angela; Ruiz-Sala, Pedro; Garcia-Villoria, Judit; Carling, Rachel; Bierau, Jörgen; Ruijter, George J G; Casado, Mercedes; Ormazabal, Aida; Artuch, Rafael.
Afiliación
  • Oliva C; Biochemistry and Molecular Genetics Department, 571524 Hospital Clínic de Barcelona , Barcelona, Spain.
  • Arias A; Clinical Biochemistry Department, 16512 Institut de Recerca Sant Joan de Déu , Barcelona, Spain.
  • Ruiz-Sala P; Centre for Biomedical Network Research on Rare Diseases (CIBERER), Instituto de Salud Carlos III, Madrid, Spain.
  • Garcia-Villoria J; Centro de Diagnóstico de Enfermedades Moleculares, Universidad Autónoma de Madrid, IdIPAZ, Madrid, Spain.
  • Carling R; Biochemistry and Molecular Genetics Department, 571524 Hospital Clínic de Barcelona , Barcelona, Spain.
  • Bierau J; Centre for Biomedical Network Research on Rare Diseases (CIBERER), Instituto de Salud Carlos III, Madrid, Spain.
  • Ruijter GJG; Institut d'Investigacions Biomèdiques August Pi i Sunyer (IDIBAPS), Barcelona, Spain.
  • Casado M; Department of Biochemical Sciences, 8945 Synnovis, Guy's & St Thomas' NHSFT , London, UK.
  • Ormazabal A; Department of Clinical Genetics, 570888 Maastricht University Medical Center , Maastricht, The Netherlands.
  • Artuch R; Department of Clinical Genetics, Erasmus Medical Center, Rotterdam, The Netherlands.
Clin Chem Lab Med ; 62(10): 1991-2000, 2024 Sep 25.
Article en En | MEDLINE | ID: mdl-38456798
ABSTRACT

OBJECTIVES:

Early diagnosis of inborn errors of metabolism (IEM) is crucial to ensure early detection of conditions which are treatable. This study reports on targeted metabolomic procedures for the diagnosis of IEM of amino acids, acylcarnitines, creatine/guanidinoacetate, purines/pyrimidines and oligosaccharides, and describes its validation through external quality assessment schemes (EQA).

METHODS:

Analysis was performed on a Waters ACQUITY UPLC H-class system coupled to a Waters Xevo triple-quadrupole (TQD) mass spectrometer, operating in both positive and negative electrospray ionization mode. Chromatographic separation was performed on a CORTECS C18 column (2.1 × 150, 1.6 µm). Data were collected by multiple reaction monitoring.

RESULTS:

The internal and EQA results were generally adequate, with a few exceptions. We calculated the relative measurement error (RME) and only a few metabolites displayed a RME higher than 30 % (asparagine and some acylcarnitine species). For oligosaccharides, semi-quantitative analysis of an educational panel clearly identified the 8 different diseases included.

CONCLUSIONS:

Overall, we have validated our analytical system through an external quality control assessment. This validation will contribute to harmonization between laboratories, thus improving identification and management of patients with IEM.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Espectrometría de Masas en Tándem / Errores Innatos del Metabolismo Límite: Humans Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2024 Tipo del documento: Article País de afiliación: España

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Espectrometría de Masas en Tándem / Errores Innatos del Metabolismo Límite: Humans Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2024 Tipo del documento: Article País de afiliación: España