Percutaneous Lung Biopsies With Robotic Systems: A Systematic Review of Available Clinical Solutions.
Bodard, Sylvain; Guinebert, Sylvain; Petre, Elena N; Alexander, Erica; Marinelli, Brett; Sarkar, Debkumar; Cornelis, Francois H.
Afiliación
  • Bodard S; Department of Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
  • Guinebert S; Department of Radiology, University of Paris Cité, Necker Hospital, Paris, France.
  • Petre EN; Laboratoire d'Imagerie Biomédicale, Sorbonne University, CNRS UMR 7371, INSERM U 1146, Paris, France.
  • Alexander E; Department of Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
  • Marinelli B; Department of Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
  • Sarkar D; Department of Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
  • Cornelis FH; Department of Radiology, Memorial Sloan Kettering Cancer Center, New York, NY, USA.
Can Assoc Radiol J ; 75(4): 907-920, 2024 Nov.
Article en En | MEDLINE | ID: mdl-38581355
ABSTRACT

Objectives:

This systematic review aims to assess existing research concerning the use of robotic systems to execute percutaneous lung biopsy.

Methods:

A systematic review was performed and identified 4 studies involving robotic systems used for lung biopsy. Outcomes assessed were operation time, radiation dose to patients and operators, technical success rate, diagnostic yield, and complication rate.

Results:

One hundred and thirteen robot-guided percutaneous lung biopsies were included. Technical success and diagnostic yield were close to 100%, comparable to manual procedures. Technical accuracy, illustrated by needle positioning, showed less frequent needle adjustments in robotic guidance than in manual guidance (P < .001) 2.7 ± 2.6 (range 1-4) versus 6 ± 4 (range 2-12). Procedure time ranged from comparable to reduced by 35% on average (20.1 ± 11.3 minutes vs 31.4 ± 10.2 minutes, P = .001) compared to manual procedures. Patient irradiation ranged from comparable to reduced by an average of 40% (324 ± 114.5 mGy vs 541.2 ± 446.8 mGy, P = .001). There was no significant difference in reported complications between manual biopsy and biopsies that utilized robotic guidance.

Conclusion:

Robotic systems demonstrate promising results for percutaneous lung biopsy. These devices provide adequate accuracy in probe placement and could both reduce procedural duration and mitigate radiation exposure to patients and practitioners. However, this review underscores the need for larger, controlled trials to validate and extend these findings.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Biopsia Guiada por Imagen / Procedimientos Quirúrgicos Robotizados / Pulmón Límite: Humans Idioma: En Revista: Can Assoc Radiol J Asunto de la revista: RADIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Biopsia Guiada por Imagen / Procedimientos Quirúrgicos Robotizados / Pulmón Límite: Humans Idioma: En Revista: Can Assoc Radiol J Asunto de la revista: RADIOLOGIA Año: 2024 Tipo del documento: Article País de afiliación: Estados Unidos