Your browser doesn't support javascript.
loading
Analytical validation of the Mindray CL1200i analyzer high sensitivity cardiac troponin I assay: MERITnI study.
Fabre-Estremera, Blanca; Schulz, Karen; Ladd, Alanna; Sexter, Anne; Apple, Fred S.
Afiliación
  • Fabre-Estremera B; Department of Laboratory Medicine, 16268 La Paz University Hospital , Madrid, Spain.
  • Schulz K; Department of Preventive Medicine and Public Health, Universidad Autónoma de Madrid, Madrid, Spain.
  • Ladd A; 20298 Hennepin Healthcare Research Institute , Minneapolis, MN, USA.
  • Sexter A; 20298 Hennepin Healthcare Research Institute , Minneapolis, MN, USA.
  • Apple FS; 20298 Hennepin Healthcare Research Institute , Minneapolis, MN, USA.
Clin Chem Lab Med ; 2024 May 28.
Article en En | MEDLINE | ID: mdl-38801528
ABSTRACT

OBJECTIVES:

This study performed an analytical validation study of the Mindray high-sensitivity cardiac troponin I (hs-cTnI) assay addressing limit of blank (LoB), limit of detection (LoD), precision, linearity, analytical specificity and sex-specific 99th percentile upper reference limits.

METHODS:

LoB, LoD, precision, linearity and analytical specificity were studied according to Clinical and Laboratory Standards Institute. We used one reagent lot and one CL1200i analyzer. Skeletal troponin I and T, cardiac troponin T, troponin C, actin, tropomyosin, myosin light chain, myoglobin and creatine kinase (CK-MB) were studied for cross-reactivity. Interference with biotin was examined. Lithium heparin samples (one freeze thaw cycle) from healthy males and females were measured to determine the 99th percentiles by using the non-parametric method. Analyses were performed before and after excluding subjects with clinical conditions and/or increased surrogate biomarkers.

RESULTS:

The Mindray hs-cTnI assay met criteria to be considered as a hs-cTn assay. LoB and LoD was <0.1 ng/L and 0.1 ng/L, respectively. Repeatability had a coefficient of variation 1.2-3.8 %, and within-laboratory imprecision 1.7-5.0 %. The measuring interval ranged from 1.1 to 28,180 ng/L. The analytical specificity was clinically acceptable for the interferents studied. After exclusions, the 99th percentile URLs obtained were 10 ng/L overall, 5 ng/L for females and 12 ng/L for males.

CONCLUSIONS:

Analytical observations of the Mindray hs-cTnI assay demonstrated excellent LoB, LoD, precision, linearity and analytical specificity, that were in alignment with the manufacturer's claims and regulatory guidelines for hs-cTnI. The assay is suitable for clinical investigation for patient-oriented studies.
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2024 Tipo del documento: Article País de afiliación: España

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Idioma: En Revista: Clin Chem Lab Med Asunto de la revista: QUIMICA CLINICA / TECNICAS E PROCEDIMENTOS DE LABORATORIO Año: 2024 Tipo del documento: Article País de afiliación: España