Free-breathing 3D whole-heart joint T1/T2 mapping and water/fat imaging at 0.55 T.
Si, Dongyue; Crabb, Michael G; Kunze, Karl P; Littlewood, Simon J; Prieto, Claudia; Botnar, René M.
Afiliación
  • Si D; School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • Crabb MG; School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • Kunze KP; School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • Littlewood SJ; MR Research Collaborations, Siemens Healthcare Limited, Camberley, UK.
  • Prieto C; School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
  • Botnar RM; School of Biomedical Engineering and Imaging Sciences, King's College London, London, UK.
Magn Reson Med ; 92(4): 1511-1524, 2024 Oct.
Article en En | MEDLINE | ID: mdl-38872384
ABSTRACT

PURPOSE:

To develop and validate a highly efficient motion compensated free-breathing isotropic resolution 3D whole-heart joint T1/T2 mapping sequence with anatomical water/fat imaging at 0.55 T.

METHODS:

The proposed sequence takes advantage of shorter T1 at 0.55 T to acquire three interleaved water/fat volumes with inversion-recovery preparation, no preparation, and T2 preparation, respectively. Image navigators were used to facilitate nonrigid motion-compensated image reconstruction. T1 and T2 maps were jointly calculated by a dictionary matching method. Validations were performed with simulation, phantom, and in vivo experiments on 10 healthy volunteers and 1 patient. The performance of the proposed sequence was compared with conventional 2D mapping sequences including modified Look-Locker inversion recovery and T2-prepared balanced steady-SSFP sequence.

RESULTS:

The proposed sequence has a good T1 and T2 encoding sensitivity in simulation, and excellent agreement with spin-echo reference T1 and T2 values was observed in a standardized T1/T2 phantom (R2 = 0.99). In vivo experiments provided good-quality co-registered 3D whole-heart T1 and T2 maps with 2-mm isotropic resolution in a short scan time of about 7 min. For healthy volunteers, left-ventricle T1 mean and SD measured by the proposed sequence were both comparable with those of modified Look-Locker inversion recovery (640 ± 35 vs. 630 ± 25 ms [p = 0.44] and 49.9 ± 9.3 vs. 54.4 ± 20.5 ms [p = 0.42]), whereas left-ventricle T2 mean and SD measured by the proposed sequence were both slightly lower than those of T2-prepared balanced SSFP (53.8 ± 5.5 vs. 58.6 ± 3.3 ms [p < 0.01] and 5.2 ± 0.9 vs. 6.1 ± 0.8 ms [p = 0.03]). Myocardial T1 and T2 in the patient measured by the proposed sequence were in good agreement with conventional 2D sequences and late gadolinium enhancement.

CONCLUSION:

The proposed sequence simultaneously acquires 3D whole-heart T1 and T2 mapping with anatomical water/fat imaging at 0.55 T in a fast and efficient 7-min scan. Further investigation in patients with cardiovascular disease is now warranted.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Tejido Adiposo / Fantasmas de Imagen / Imagenología Tridimensional Límite: Adult / Female / Humans / Male Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2024 Tipo del documento: Article

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Tejido Adiposo / Fantasmas de Imagen / Imagenología Tridimensional Límite: Adult / Female / Humans / Male Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 2024 Tipo del documento: Article