Development and validation of an ultra-performance liquid chromatography-tandem mass spectrometry method to quantify the small molecule inhibitors adagrasib, alectinib, brigatinib, capmatinib, crizotinib, lorlatinib, selpercatinib, and sotorasib in human plasma.
Hollander, Esther M; Zimmerman, Nachel M C; Piet, Berber; van den Heuvel, Michel M; Burger, David M; Te Brake, Lindsey H M; Ter Heine, Rob.
Afiliación
  • Hollander EM; Department of Pharmacy and Research Institute for Medical Innovation, Radboud University Medical Center, Nijmegen, The Netherlands.
  • Zimmerman NMC; Department of Pharmacy and Research Institute for Medical Innovation, Radboud University Medical Center, Nijmegen, The Netherlands.
  • Piet B; Department of Pulmonology, Radboud University Medical Center, Nijmegen, The Netherlands.
  • van den Heuvel MM; Department of Pulmonology, Radboud University Medical Center, Nijmegen, The Netherlands.
  • Burger DM; Department of Pharmacy and Research Institute for Medical Innovation, Radboud University Medical Center, Nijmegen, The Netherlands.
  • Te Brake LHM; Department of Pharmacy and Research Institute for Medical Innovation, Radboud University Medical Center, Nijmegen, The Netherlands.
  • Ter Heine R; Department of Pharmacy and Research Institute for Medical Innovation, Radboud University Medical Center, Nijmegen, The Netherlands.
Biomed Chromatogr ; 38(10): e5986, 2024 Oct.
Article en En | MEDLINE | ID: mdl-39136165
ABSTRACT
Small molecule inhibitors (SMIs) are increasingly being used in the treatment of non-small cell lung cancer. To support pharmacokinetic research and clinical treatment monitoring, our aim was to develop and validate an ultra-performance liquid chromatography-mass spectrometry (UPLC-MS/MS) assay for quantification of eight SMIs adagrasib, alectinib, brigatinib, capmatinib, crizotinib, lorlatinib, selpercatinib, and sotorasib. Development of the UPLC-MS/MS assay was done by trying different columns and eluents to optimize peak shape. The assay was validated based on guidelines of the European Medicines Agency. Chromatographic separation was performed with a gradient elution using ammonium formate in water and methanol. Detection was performed using a triple quadrupole tandem mass spectrometer with electrospray ionization. Validation was performed in a range of 10-2500 µg/L for lorlatinib, 25-6250 µg/L for alectinib and crizotinib, 25-10,000 µg/L for capmatinib and selpercatinib, 50-12,500 µg/L for brigatinib, and 100-25,000 µg/L for adagrasib and sotorasib. Imprecision was <8.88% and inaccuracy was <12.5% for all compounds. Seven out of eight compounds were stable for 96 h at room temperature. Sotorasib was stable for 8 h at room temperature. A sensitive and reliable method has been developed to quantify eight SMIs with a single assay, enhancing efficacy and safety of targeted therapies.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Espectrometría de Masas en Tándem Límite: Humans Idioma: En Revista: Biomed Chromatogr Año: 2024 Tipo del documento: Article País de afiliación: Países Bajos

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Espectrometría de Masas en Tándem Límite: Humans Idioma: En Revista: Biomed Chromatogr Año: 2024 Tipo del documento: Article País de afiliación: Países Bajos