Detection of spontaneous breathing during an apnea test in a patient with suspected brain death using electrical impedance tomography: a case report.
Chen, Rongqing; Lovas, András; Bakos, Péter; Molnár, Tamás; Hawchar, Fatime; Benyó, Balázs; Zhao, Zhanqi; Chase, J Geoffrey; Rupitsch, Stefan J; Moeller, Knut.
Afiliación
  • Chen R; Institute of Technical Medicine, Hochschule Furtwangen, Jakob-Kienzle-Str. 17, Villingen-Schwenningen, 78054, Germany. rongqing.chen@hs-furtwangen.de.
  • Lovas A; Department of Microsystems Engineering (IMTEK), Faculty of Engineering, University of Freiburg, Georges-Köhler-Allee 101, Freiburg, 79110, Germany. rongqing.chen@hs-furtwangen.de.
  • Bakos P; Department of Anaesthesiology and Intensive Therapy, Kiskunhalas Semmelweis Hospital, Teaching Hospital of the University of Szeged, Dr. Monszpart László u. 1, Kiskunhalas, 6400, Hungary.
  • Molnár T; Department of Anesthesiology and Intensive Therapy, Csolnoky Ferenc Hospital, Kórház u. 1, Veszprém, 8200, Hungary.
  • Hawchar F; Department of Cardiology, State Hospital for Cardiology, Gyógy tér 2, Balatonfüred, 8230, Hungary.
  • Benyó B; Department of Anesthesiology and Intensive Therapy, University of Szeged, Semmelweis u. 6, Szeged, 6725, Hungary.
  • Zhao Z; Department of Anesthesiology and Intensive Therapy, Budapesti Dr. Manninger Jeno Traumatology Center, Fiumei út 17, Budapest, 1080, Hungary.
  • Chase JG; Department of Control Engineering and Information Technology, Faculty of Electrical Engineering and Information Technology, Budapest University of Technology and Economics, Muegyetem rkp. 3, Budapest, 1111, Hungary.
  • Rupitsch SJ; Institute of Technical Medicine, Hochschule Furtwangen, Jakob-Kienzle-Str. 17, Villingen-Schwenningen, 78054, Germany.
  • Moeller K; Department of Mechanical Engineering, University of Canterbury, 69 Creyke Road, Christchurch, 8041, New Zealand.
BMC Pulm Med ; 24(1): 454, 2024 Sep 16.
Article en En | MEDLINE | ID: mdl-39285376
ABSTRACT

INTRODUCTION:

The apnea test (AT) is a crucial procedure in determining brain death (BD), with detection of spontaneous breathing efforts serving as a key criterion. Numerous national statutes mandate complete disconnection of the patient from the ventilator during the procedure to open the airway directly to the atmosphere. These regulations mandate visual observation as an exclusive option for detecting breathing efforts. However, reliance on visual observation alone can pose challenges in identifying subtle respiratory movements. CASE PRESENTATION This case report presents a 55-year-old morbidly obese male patient with suspected BD due to cerebral hemorrhage undergoing an AT. The AT was performed with continuous electrical impedance tomography (EIT) monitoring. Upon detection of spontaneous breathing movements by both visual observation and EIT, the AT was aborted, and the patient was reconnected to the ventilator. EIT indicated a shift in ventilation distribution from the ventral to the dorsal regions, indicating the presence of spontaneous breathing efforts. EIT results also suggested the patient experienced a slow but transient initial recovery phase, likely due to atelectasis induced by morbid obesity, before returning to a steady state of ventilatory support.

CONCLUSION:

The findings suggest EIT could enhance the sensitivity and accuracy of detecting spontaneous breathing efforts, providing additional insights into the respiratory status of patients during the AT.
Asunto(s)
Palabras clave

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Apnea / Obesidad Mórbida / Muerte Encefálica / Tomografía / Impedancia Eléctrica Límite: Humans / Male / Middle aged Idioma: En Revista: BMC Pulm Med Año: 2024 Tipo del documento: Article País de afiliación: Alemania

Texto completo: 1 Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Apnea / Obesidad Mórbida / Muerte Encefálica / Tomografía / Impedancia Eléctrica Límite: Humans / Male / Middle aged Idioma: En Revista: BMC Pulm Med Año: 2024 Tipo del documento: Article País de afiliación: Alemania