In-plane motion correction for MR spectroscopic imaging.
Haupt, C I; Kiefer, A P; Maudsley, A A.
Afiliación
  • Haupt CI; Department of Radiology, University of California San Francisco, USA.
Magn Reson Med ; 39(5): 749-53, 1998 May.
Article en En | MEDLINE | ID: mdl-9581606
ABSTRACT
A motion-detection method is described that is specifically suited for MR spectroscopic imaging (MRSI) studies. Information on in-plane rotation and translation of the subject was obtained using external spatial reference markers that are uniquely identified via their chemical shift. The marker locations were obtained directly from the acquired data at each encoding step, and no additional data acquisition was required. This method was applied to brain 1H MRSI studies that include subcutaneous lipid signals, which otherwise result in enhanced sensitivity to subject motion.
Asunto(s)
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Encéfalo / Imagen por Resonancia Magnética / Espectroscopía de Resonancia Magnética / Artefactos Límite: Female / Humans / Male Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 1998 Tipo del documento: Article País de afiliación: Estados Unidos
Buscar en Google
Colección: 01-internacional Banco de datos: MEDLINE Asunto principal: Encéfalo / Imagen por Resonancia Magnética / Espectroscopía de Resonancia Magnética / Artefactos Límite: Female / Humans / Male Idioma: En Revista: Magn Reson Med Asunto de la revista: DIAGNOSTICO POR IMAGEM Año: 1998 Tipo del documento: Article País de afiliación: Estados Unidos