Your browser doesn't support javascript.
loading
In vivo dosimetry for estimation of effective doses in multislice CT coronary angiography.
de Denaro, M; Bregant, P; Severgnini, M; de Guarrini, F.
Afiliação
  • de Denaro M; Department of Medical Physics, A.O.U. "Ospedali Riuniti" Trieste, Italy. mario.dedenaro@aots.sanita.fvg.it
Med Phys ; 34(10): 3705-10, 2007 Oct.
Article em En | MEDLINE | ID: mdl-17985615
ABSTRACT
In vivo dosimetry represents a technique that has been widely employed to evaluate the dose to the patient mainly in radiotherapy. Considering the increment in dose to the population due to new high-dose multislice CT examinations, such as coronary angiography, it is becoming important to more accurately know the dose to the patient. The desire to know patient dose extends even to radiological examinations. Thermoluminescent dosimeters are considered the gold standard for in vivo dosimetry, but their use is time consuming. A rapid, less labor-intensive method has been developed to perform in vivo dosimetry using radiochromic film positioned next to the patient's skin. Multislice CT scanners allow the estimation of the effective dose to the patient from the dose length product (DLP) parameter, the value of which is displayed on the acquisition console, simply multiplying the DLP by published conversion factors. The method represents only an approximation based on standard size circular phantoms and neglects the actual size of the patient. More accurate evaluations can be carried out using software-based Monte Carlo simulations. However, these methods do not consider possible dose reduction techniques, such as automatic tube-current modulation. For 22 patients effective doses measured by in vivo dosimetry and calculated by software were compared. The technique of using in vivo dosimetry measured with radiochromic film appears a promising procedure for improving the assessment of the effective dose to the patient.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Tomografia Computadorizada por Raios X Tipo de estudo: Health_economic_evaluation Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 2007 Tipo de documento: Article País de afiliação: Itália
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Radiometria / Tomografia Computadorizada por Raios X Tipo de estudo: Health_economic_evaluation Limite: Humans Idioma: En Revista: Med Phys Ano de publicação: 2007 Tipo de documento: Article País de afiliação: Itália