Comparison of torque measurements and clinical handling of various surgical motors.
Neugebauer, Jörg; Scheer, Martin; Mischkowski, Robert A; An, Sang-Ho; Karapetian, Viktor E; Toutenburg, Helge; Zoeller, Joachim E.
Afiliação
  • Neugebauer J; Interdisciplinary Department of Oral Surgery and Implantology, University of Cologne, Köln, Germany. Joerg.neugebauer@uk-koeln.de
Int J Oral Maxillofac Implants ; 24(3): 469-76, 2009.
Article em En | MEDLINE | ID: mdl-19587869
ABSTRACT

PURPOSE:

Modern implant dentistry requires the application of torque during various treatment steps. This study investigated seven different surgical motors for the accuracy of the applied torque and their reliability. MATERIALS AND

METHODS:

The following surgical motors were evaluated Chiropro 980 (Bienair), INTRAsurg 300 and INTRAsurg 500 (KaVo), Osseocare (Nobel Biocare), Surgic XT (NSK), Elcomed SA-200 C (W and H), and Osseo System (XO Dentalcare). The torque was measured during typical surgical and prosthetic procedures using a special load transfer mechanism for a torque gauge. For each setting, 30 measurements were made and means were calculated.

RESULTS:

The highest percentage shortfall was 20.5% at a set torque of 11.4 Ncm (absolute deviation of -2.4 Ncm). The highest percentage by which a torque was exceeded was 54.6% (absolute deviation of 5.5 Ncm). The lowest value for absolute shortfall was found to be -5.6 Ncm at a set torque of 45 Ncm. The highest absolute exceeded was 15 Ncm at a set torque of 40 Ncm. Potentially problematic torque values were identified in the low-torque-value setting, as the implant position may be changed if a machine driver applies excessive torque to the first-stage healing screw. In addition, in the indication of immediate loading in the high-set-torque group, torque values above the critical value of 50 Ncm may be unwittingly applied while working with a set torque of 40 Ncm.

CONCLUSION:

For most of the clinically relevant torque settings, precise values were measured, although a few devices delivered potentially problematic torque values for some of the indications.
Assuntos
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Implantação Dentária Endóssea / Análise do Estresse Dentário Limite: Humans Idioma: En Revista: Int J Oral Maxillofac Implants Assunto da revista: ODONTOLOGIA Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Alemanha
Buscar no Google
Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Implantação Dentária Endóssea / Análise do Estresse Dentário Limite: Humans Idioma: En Revista: Int J Oral Maxillofac Implants Assunto da revista: ODONTOLOGIA Ano de publicação: 2009 Tipo de documento: Article País de afiliação: Alemanha