MRI-guided vascular access with an active visualization needle.
Saikus, Christina E; Ratnayaka, Kanishka; Barbash, Israel M; Colyer, Jessica H; Kocaturk, Ozgur; Faranesh, Anthony Z; Lederman, Robert J.
Afiliação
  • Saikus CE; Cardiovascular and Pulmonary Branch, Division of Intramural Research, National Heart, Lung, and Blood Institute, National Institutes of Health, Bethesda, Maryland 20892-1061, USA.
J Magn Reson Imaging ; 34(5): 1159-66, 2011 Nov.
Article em En | MEDLINE | ID: mdl-22006552
ABSTRACT

PURPOSE:

To develop an approach to vascular access under magnetic resonance imaging (MRI), as a component of comprehensive MRI-guided cardiovascular catheterization and intervention. MATERIALS AND

METHODS:

We attempted jugular vein access in healthy pigs as a model of "difficult" vascular access. Procedures were performed under real-time MRI guidance using reduced field of view imaging. We developed an "active" MRI antenna-needle having an open-lumen, distinct tip appearance and indicators of depth and trajectory in order to enhance MRI visibility during the procedure. We compared performance of the active needle against an unmodified commercial passively visualized needle, measured by procedure success among operators with different levels of experience.

RESULTS:

MRI-guided central vein access was feasible using both the active needle and the unmodified passive needle. The active needle required less time (88 vs. 244 sec, P = 0.022) and fewer needle passes (4.5 vs. 9.1, P = 0.028), irrespective of operator experience.

CONCLUSION:

MRI-guided access to central veins is feasible in our animal model. When image guidance is necessary for vascular access, performing this component under MRI will allow wholly MRI-guided catheterization procedures that do not require adjunctive imaging facilities such as x-ray or ultrasound. The active needle design showed enhanced visibility, as expected. These capabilities may permit more complex catheter-based cardiovascular interventional procedures enabled by enhanced image guidance.
Assuntos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cateterismo Venoso Central / Diagnóstico por Imagem / Imageamento por Ressonância Magnética / Veias Jugulares Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Animals Idioma: En Revista: J Magn Reson Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Estados Unidos

Texto completo: 1 Coleções: 01-internacional Base de dados: MEDLINE Assunto principal: Cateterismo Venoso Central / Diagnóstico por Imagem / Imageamento por Ressonância Magnética / Veias Jugulares Tipo de estudo: Diagnostic_studies / Prognostic_studies Limite: Animals Idioma: En Revista: J Magn Reson Imaging Assunto da revista: DIAGNOSTICO POR IMAGEM Ano de publicação: 2011 Tipo de documento: Article País de afiliação: Estados Unidos